CARE OF THE INSTRUMENT
7.12
Revised 10/02/2012
1.
Open the Fume Control System Access Door (Figure
7-W).
Figure 7-W
2.
Release the latch securing the fume filter access door
(Figure 7-X) and open the access door.
Figure 7-X
3.
Unlock the latch securing the fume filter cover (Figure
7-Y) and lift the cover up.
Figure 7-Y
4.
Slide the fume filter out of the instrument (Figure 7-Z).
Figure 7-Z
5.
Install a new fume filter in the instrument. Lower the
fume filter cover and secure the cover latch.
6.
Close the fume filter access door and secure the door
latch (Figure 7-AA).
Figure 7-AA
7.
Close the Fume Control System Access Door.