Revised 10/02/2012
4.7
4.
Place the basket or magazine set directly into the
Loading Station retort (Figure 4-J). Close the Loading
Station Access Door.
Figure 4-J
CAUTION: The Loading Station retort has slots (
..
1
..,
Figure 4-K) positioned at the front and rear of the
retort. Make sure the tabs of the basket handle engage
these slots when the basket or magazine set is placed
in the retort. If the basket is not positioned properly,
the Transfer Arm will not be able to grasp it firmly.
Figure 4-K
5.
Close the Loading Station Access Door.
•
If
REAGENT CONTROL is set to CASSETTES,
the CASSETTE COUNT data entry dialog dis-
plays (Figure 4-L). Use the keypad to enter the
number of cassettes contained in the basket or
magazine set to be processed in the
INPUT
NUMBER OF CASSETTES text box.
Figure 4-L
NOTE: If needed, use the backspace key to correct
errors when entering the number of cassettes. Touch the
CANCEL key to delete
all characters
from the
INPUT
NUMBER OF CASSETTES text box. If
no value
is
entered in the
INPUT NUMBER OF CASSETTES box, the
system will default to 40 cassettes.
•
If
REAGENT CONTROL is set to RUN or DAY,
the system counts each basket or magazine set
as a run and a day is automatically accounted
for; no user entry is required.
•
Select the appropriate user-defined item (loca-
tion, department, or doctor’s name) for which the
cassettes are being processed from the
SELECT
ITEM menu.
NOTE: If the USER DEFINED FILE is not enabled (see
Configuring System Settings
on page 3.2 for details), the
SELECT ITEM menu will not be available.
6.
When the necessary information has been entered,
touch the
ENTER button to complete the processing
cycle setup. The CASSETTE COUNT dialog closes.
7.
The instrument begins the processing cycle.
•
If the preprocessing option is chosen, the basket
or magazine will remain in the loading station for
the predetermined time before transferring to
Station #1.
•
Once the basket or magazine set has been trans-
ferred from the Loading Station to Retort #1, an-
other basket or magazine set may be loaded into
the instrument for processing. The start time for
the next basket or magazine set depends on the
Содержание Tissue-Tek Xpress x120
Страница 1: ...Operating Operating Manual Manual Rapid Tissue Processor Tissue Tek Xpress x120 ...
Страница 3: ......
Страница 20: ...INTRODUCTION 1 14 Revised 10 02 2012 ...
Страница 24: ...INSTALLATION 2 4 Revised 10 02 2012 ...
Страница 54: ...RECOVERY PROCESS 5 4 Revised 10 02 2012 ...
Страница 82: ...A 8 Revised 10 02 2012 ...