Revised 10/02/2012
7.11
Preheating Chamber Sight Tube
Cleaning
The Preheating Chamber sight tubes should be
cleaned weekly to ensure reagent in Retorts #1 and #2 is
filled and maintained at the appropriate levels for proper
processing of tissue specimens, and to guard against
reagent overflow.
WARNING: MAY CONTAIN BIOHAZARDOUS WASTE.
DISPOSE OF CONTENTS IN ACCORDANCE WITH
LOCAL REGULATIONS.
To clean the preheating chamber sight tubes:
CAUTION: Ensure the system shutdown has complet-
ed prior to opening the valve.
1.
Open the preheating chamber access door on the
left-hand side of the instrument.
NOTE: Prior to opening the valves, place a small contain-
er under the valves to catch any reagent within the site
tubes.
2.
Turn the sight tube access valve for the Retort #1
preheating chamber counterclockwise to open the
valve (Figure 7-U).
Figure 7-U
3.
Insert the sight tube cleaning brush (see
“Section 6,
Consumables and Accessories”
for details) into the
sight tube, and scrub up and down several times to re-
move debris from the wall of the sight tube (Figure 7-V).
4.
After cleaning, turn the sight tube access valve for
Retort #1 preheating chamber clockwise to close the
valve.
5.
Repeat steps 2 through 4 for the Retort #2 sight tube.
6.
Close the preheating chamber access door.
Figure 7-V
Monthly Maintenance
Fume Control System Filter Re-
placement
WARNING: AN ACTIVATED CARBON FILTER MUST
BE INSTALLED IN THE INSTRUMENT AT ALL TIMES
TO ENSURE PROPER FILTERING OF FUMES
GENERATED DURING PROCESSING.
The fume control system serves to reduce acetone
emissions generated by the instrument to acceptable
levels. To ensure proper filtering and fume control, replace
the fume control system filter monthly.
To replace the fume control system activated carbon
filter:
Содержание Tissue-Tek Xpress x120
Страница 1: ...Operating Operating Manual Manual Rapid Tissue Processor Tissue Tek Xpress x120 ...
Страница 3: ......
Страница 20: ...INTRODUCTION 1 14 Revised 10 02 2012 ...
Страница 24: ...INSTALLATION 2 4 Revised 10 02 2012 ...
Страница 54: ...RECOVERY PROCESS 5 4 Revised 10 02 2012 ...
Страница 82: ...A 8 Revised 10 02 2012 ...