INTRODUCTION
1.4
Revised 10/02/2012
Unloading Indicator
..
4
.. — provides an indication of the
status of the Unloading Station access door:
Light Off – indicates “access denied” to the Unloading
Station access door; the door should not be opened. The
Unloading Indicator is “off” whenever the Unloading
Station is empty or when the Transferring Arm is in
motion.
Light On – indicates “access permitted” to the Unloading
Station access door; the door can be opened. The
Unloading Indicator is “on” when there is a basket in the
Unloading Station and the Transferring Arm is stationary.
Flashing – alerts the operator to remove the basket(s)
from the unloading station or the access door is open
and needs to be closed immediately.
Audible Alarm (not shown) – indicates the following:
•
A tone sounds at the completion of a processing cycle
(when a basket has arrived at the Unloading Station).
Tone selection (from seven available tones), volume
selection (high, middle, low), and pattern selection
(continuous, intermittent, 30 seconds only) are user
selectable (see “
Configuring System Settings
” on page
3.2 for details).
•
A continuous tone sounds if an abnormal condition is
detected during processing. The alarm condition must
be acknowledged and/or cleared by the operator to
cancel the alarm. Tone selection (from seven availa-
ble tones) and volume selection (high, mid, low) are
user selectable (see “
Configuring System Settings
” on
page 3.2 for details).
Retort Access Door
..
5
.. — provides access to the
Microwave and Vacuum Stations for purposes of cleaning
and/or maintenance. Access to all four retorts is permit-
ted/denied, under software control, when the instrument is
processing specimens.
Retort #1 / #2 Reagent Cabinet Access Door
..
6
.. —
provides access to retort #1 / #2 Reagent Cabinet for the
purposes of replacing reagents.
Retort #3 / #4 Reagent Cabinet Access Door
..
7
.. —
provides access to the paraffin oven for the purposes of
replacing reagents.
Control Panel
..
8
.. — comprises the controls and
indicators necessary to program and operate the instru-
ment, and to monitor the instrument during processing
(refer to
“Control Panel (Figure 1-O)”
on page 1.12 for
details).
Fume Control System Access Door
..
9
.. — provides
access to the fume control system for purposes of
replacing the fume filter.
Overflow Bottle Access Door
..
10
.. — provides access
to retorts #3 / #4 overflow bottles for purposes of cleaning
and servicing.
Leveling Feet/Casters
..
11
.. — four casters are provided
to allow for easy repositioning of the instrument. Adjusta-
ble leveling feet associated with each caster facilitate
proper leveling of the instrument.
Cabinet Rear
(Figure 1-B)
The power cord connector and external interface are
located at the rear of the back cabinet.
Figure 1-B
Power Cord Connector
..
12
.. — accepts the instrument
end of the power cord; opposite end of the power cord
may be connected to facility power.
Содержание Tissue-Tek Xpress x120
Страница 1: ...Operating Operating Manual Manual Rapid Tissue Processor Tissue Tek Xpress x120 ...
Страница 3: ......
Страница 20: ...INTRODUCTION 1 14 Revised 10 02 2012 ...
Страница 24: ...INSTALLATION 2 4 Revised 10 02 2012 ...
Страница 54: ...RECOVERY PROCESS 5 4 Revised 10 02 2012 ...
Страница 82: ...A 8 Revised 10 02 2012 ...