Revised 10/02/2012
7.5
NOTE: Solely shaking the reagent bottle will not thorough-
ly mix the solution. The bottle must be inverted.
6.
Remove the upper caps from the new reagent
containers, and install the caps on the used reagent
containers. Discard the used reagent containers in
accordance with local regulations for biohazardous
waste.
7.
Remove the foil seal from each reagent container.
8.
Place the new reagent containers in the reagent
cabinet in the appropriate positions.
9.
Press down on each reagent container locking lever
until it engages the “LOCKED” position, coupling the
reagent containers with the reagent supply system
(Figure 7-E).
Figure 7-E
10.
Close Retort #1/#2 Reagent Cabinet Access Door.
Retort #3 and #4 Paraffin Replacement
NOTE: Retain the shipping materials to drain used
paraffin.
WARNING: DISPOSE OF USED PARAFFIN IN
ACCORDANCE WITH LOCAL REGULATIONS FOR
BIOHAZARDOUS WASTE.
To replace paraffin:
1.
Perform a system shutdown.
2.
Open the Paraffin Oven Access Door (Figure 7-F).
Figure 7-F
3.
Remove the drip pan from the paraffin oven (Figure 7-
G). If necessary, remove hardened paraffin residue
from the drip pan using the paraffin scraper (see
“Section 6, Consumables and Accessories”
for de-
tails). Replace the drip pan on the paraffin oven.
Figure 7-G
4.
Press in on the paraffin oven to unlatch, then slowly
slide the oven out of the cabinet.
Содержание Tissue-Tek Xpress x120
Страница 1: ...Operating Operating Manual Manual Rapid Tissue Processor Tissue Tek Xpress x120 ...
Страница 3: ......
Страница 20: ...INTRODUCTION 1 14 Revised 10 02 2012 ...
Страница 24: ...INSTALLATION 2 4 Revised 10 02 2012 ...
Страница 54: ...RECOVERY PROCESS 5 4 Revised 10 02 2012 ...
Страница 82: ...A 8 Revised 10 02 2012 ...