blink  signalling  an  error.  The  calibration  process  will  take  a  few 
seconds.  If  the  procedure  completed  successfully  the  CAL  led 
will turn off, otherwise it will blink and remain in that state. In this 
case, check the batteries in the probe head and rerun calibration. 

WARNING REGARDING ISOLATION 
 

Due  to  their  non-conductive  nature,  the  fiber  optic  cables 
provide  excellent  isolation  between  probe  head  and  receiver. 
However the probe head is floating at the same potential of the 
measured  conductor.  Thus  the  whole  body  of  the  XFVP  probe 
head  including  the  metallic  bayonet  connectors  can  be  at 
hazardous potentials. 

It is advisable to install a voltage detector/indicator on the measured line that provides visual 
and  clear  indication  about  the  presence  of  potentially  dangerous  voltages.  SAKER  offers  the 
MVD and MVDZ series of medium voltage presence detectors. 

LOW BATTERY CONDITION 

The  probe  head  will  measure  the  voltage  of  the  batteries  to  warn  the  user  in  case  of  a  low 
battery condition. The signalling threshold has been adjusted for typical alkaline AA batteries., 
for other chemistries it may not work properly. A single red led is used at the receiver to display 
this condition.  

In case of a battery low event the BATT LOW led will quickly blink one time followed by a 4 
second pause in the following way: 

BATTERY REPLACEMENT 

BR-XFVP-Rev 6. 12/22

www.saker-mv.com

CALIBRATION MUST BE PERFORMED AT LEAST ONCE 

AFTER TURNING ON THE EQUIPMENT

THE XFVP PROBE HEAD IS NOT A HANDHELD 

INSTRUMENT AND SHOULD BE CONSIDERED TO BE 

Содержание XFVP10-2

Страница 1: ...User Manual XFVP Medium Voltage Fiber Isolated Voltage Probe Revision 6 BR XFVP Rev 6 12 22 www saker mv com...

Страница 2: ...l accuracy as compared to other optical equipment and saves lots of setup time during research and testing Moreover the XFVP probe can do hot calibrations with no input disconnect which translates int...

Страница 3: ...can be floated at very high voltages please consider all clearances before making any measurements Do not open the probe head or receiver they do not contain user serviceable parts inside Fiber cable...

Страница 4: ...0 65 C Operating temperature receiver 0 70 C ESD EMI standards IEC61000 4 2 61000 4 3 CE certi cation GPSD 2001 95 EC EMC Directive 2014 30 EU RoHS2 2011 65 EU 1 Specs for XFVP20 model 2 3dB small si...

Страница 5: ...mination is needed to provide appropriate high frequency transmission of the signal If the oscilloscope does not have internal 50 termination an optional BNC terminator code XFVPTERM can be supplied I...

Страница 6: ...of IEC EN 61010 031 RECEIVER OSCILLOSCOPE CONNECTION A BNC BNC cable is supplied to connect the receiver to a measuring oscilloscope Use of a good quality cable is recommended as it could affect the m...

Страница 7: ...ead or the receiver press the TURN ON button until the ON Led lights It takes a short press to turn on the units If the probe head fails to turn on check if the batteries are dead and replace them The...

Страница 8: ...emitter It is recommended to use this feature every time the probe is not in use even for shorts periods of time Since the probe head can be set into low power mode remotely from the receiver potenti...

Страница 9: ...boots since these protect the connectors from dust and maintain the necessary alignment for proper accuracy CALIBRATION Calibration is needed to adjust the analog parameters of the fiber optic link a...

Страница 10: ...icator on the measured line that provides visual and clear indication about the presence of potentially dangerous voltages SAKER offers the MVD and MVDZ series of medium voltage presence detectors LOW...

Страница 11: ...n cleaning the exterior Dry the probe before making any voltage measurements The probe is not waterproof The fiber cables should never be disconnected The boots in the cables protect the connectors fr...

Страница 12: ...m verification Bench Multimeter of at least 5 5 digits AC and DC accuracy better than 0 1 Keysight 34450A or similar Adjustable power supply or precision voltage source Banana to BNC cable Pomona 4531...

Страница 13: ...ne batteries in the transmitter Turn on the probe head and receiver press CAL and wait 5 minutes before starting the tests OFFSET Veri cation This test can also be performed with a short at the SMB pr...

Страница 14: ...cking generator and at least 1GHz capability 1 Set the output level of the tracking generator to 5dBm 2 Set the horizontal sweep from 1MHz to 500MHz log sweep 3 Place a jump cable from output of track...

Страница 15: ...tic Emissions IEC EN 55011 A1 2011 Radiated Emissions 30 1000 MHz Class B Electromagnetic Immunity IEC EN 61000 4 2 2010 Electrostatic Discharge 4 kV contact 8 kV air IEC EN 61000 4 3 2007 RF Radiated...