INSTALLATION AND USE 

This equipment is intended to be used only by trained personnel.  

PLACING THE PROBE HEAD 

The probe head uses a SMB connector as its input. For mechanical stability and electrical 
clearance it is always recommended to install a tripod in the probe head, making sure the the 
tripod is mostly made of plastic and not metal. 

It is recommended to leave at least 

15 mm

 of air space around the probe for each kV that the 

probe is floated. Increase this distance if faster than 15kV/us slew rates in the common mode 
voltage are expected. The higher the spacing, the lower the common mode capacitance, which 
lowers the current to earth provided by the measured circuit. 

Do not use the probe for measurements on mains circuits. This probe and its attenuators are 
not CAT rated for use on these types of circuits according to definitions of IEC/EN 61010-031. 

RECEIVER-OSCILLOSCOPE CONNECTION 

A  BNC-BNC  cable  is  supplied  to  connect  the  receiver  to  a  measuring  oscilloscope.  Use  of  a 
good quality cable is recommended as it could affect the measurement. 

Set  the  oscilloscope  vertical  scale  at  an  appropriate  scale  given  the  conversion  ratio  of  the 

XFVP and attenuator (see specifications section) in question. A 50   termination is necessary 
for calibration to take place successfully. 

The preferred method of operation is as follows: 

1. Place and secure the probe head 
2. Connect the probe head to the DUT using the appropriate attenuator. 
3. Connect the receiver to the oscilloscope 
4. Turn on the probe head and receiver. 
5. Place the unit in standby mode by shortly pressing  the ON/OFF button. From this moment 

there is no need to approach the probe head, since it can be turned on from the receiver. 
Calibration also can take place while the DUT is energized. 

Ω

BR-XFVP-Rev 6. 12/22

www.saker-mv.com

MAKE SURE NO DANGEROUS VOLTAGE POTENTIALS ARE 

PRESENT WHILE INSTALLING THE PROBE HEAD IN THE DUT

Содержание XFVP10-2

Страница 1: ...User Manual XFVP Medium Voltage Fiber Isolated Voltage Probe Revision 6 BR XFVP Rev 6 12 22 www saker mv com...

Страница 2: ...l accuracy as compared to other optical equipment and saves lots of setup time during research and testing Moreover the XFVP probe can do hot calibrations with no input disconnect which translates int...

Страница 3: ...can be floated at very high voltages please consider all clearances before making any measurements Do not open the probe head or receiver they do not contain user serviceable parts inside Fiber cable...

Страница 4: ...0 65 C Operating temperature receiver 0 70 C ESD EMI standards IEC61000 4 2 61000 4 3 CE certi cation GPSD 2001 95 EC EMC Directive 2014 30 EU RoHS2 2011 65 EU 1 Specs for XFVP20 model 2 3dB small si...

Страница 5: ...mination is needed to provide appropriate high frequency transmission of the signal If the oscilloscope does not have internal 50 termination an optional BNC terminator code XFVPTERM can be supplied I...

Страница 6: ...of IEC EN 61010 031 RECEIVER OSCILLOSCOPE CONNECTION A BNC BNC cable is supplied to connect the receiver to a measuring oscilloscope Use of a good quality cable is recommended as it could affect the m...

Страница 7: ...ead or the receiver press the TURN ON button until the ON Led lights It takes a short press to turn on the units If the probe head fails to turn on check if the batteries are dead and replace them The...

Страница 8: ...emitter It is recommended to use this feature every time the probe is not in use even for shorts periods of time Since the probe head can be set into low power mode remotely from the receiver potenti...

Страница 9: ...boots since these protect the connectors from dust and maintain the necessary alignment for proper accuracy CALIBRATION Calibration is needed to adjust the analog parameters of the fiber optic link a...

Страница 10: ...icator on the measured line that provides visual and clear indication about the presence of potentially dangerous voltages SAKER offers the MVD and MVDZ series of medium voltage presence detectors LOW...

Страница 11: ...n cleaning the exterior Dry the probe before making any voltage measurements The probe is not waterproof The fiber cables should never be disconnected The boots in the cables protect the connectors fr...

Страница 12: ...m verification Bench Multimeter of at least 5 5 digits AC and DC accuracy better than 0 1 Keysight 34450A or similar Adjustable power supply or precision voltage source Banana to BNC cable Pomona 4531...

Страница 13: ...ne batteries in the transmitter Turn on the probe head and receiver press CAL and wait 5 minutes before starting the tests OFFSET Veri cation This test can also be performed with a short at the SMB pr...

Страница 14: ...cking generator and at least 1GHz capability 1 Set the output level of the tracking generator to 5dBm 2 Set the horizontal sweep from 1MHz to 500MHz log sweep 3 Place a jump cable from output of track...

Страница 15: ...tic Emissions IEC EN 55011 A1 2011 Radiated Emissions 30 1000 MHz Class B Electromagnetic Immunity IEC EN 61000 4 2 2010 Electrostatic Discharge 4 kV contact 8 kV air IEC EN 61000 4 3 2007 RF Radiated...

Отзывы: