5.CLEANING, DISINFECTION AND STERILIZATION
Before starting the cleaning and disinfection procedure, turn off the main
switch of the equipment to avoid permanent damage.
For your protection, during the cleaning and disinfection process of the
equipment use PPE such as disposable gloves and goggles.
5.1.EQUIPMENT
The cleaning and disinfection process must be performed at each patient change.
When starting the process, check for visible dirt, such as blood or saliva.
Thoroughly clean the patient's entire contact area.
For cleaning use a clean, soft cloth moistened with mild soap and then dry with a clean, soft cloth
or paper towel.
For the disinfection process of the equipment, use foam disinfectant detergent that has active
components based on didecyldimethylammonium chloride.
Apply the disinfectant detergent foam on the surface or on a clean cloth and spread it on the surface
to be treated. Respect the antimicrobial contact time specified by the manufacturer.
After application, allow to dry. Do not rinse.
Some of the removable parts that come in contact with the patient can be autoclaved. These parts
are: Ultrasound cover, Tips, Bicarbonate jet covers and Insert grip wrench.
5.1.1.Ultrasound
5.1.2.Bicarbonate Jet
Remove the Tip of the transducer, and then remove the cover by means of pressure, "do not try to
rotate", then take it for autoclaving (packed).
Unthread the handpiece of the bicarbonate jet (A) and then remove the hose (B), as it cannot be
autoclavable and take it for sterilization in an autoclave (packed).
B
A
Содержание Sonic Duo
Страница 1: ......
Страница 2: ...2 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 24: ...24 Accessories that do not accompany the product 01 07 13 04 10 02 08 14 03 09 15 05 11 06 12 ...
Страница 56: ...56 10 6 EQUIPMENT DIMENSIONS 10 6 1 Sonic Duo ...
Страница 57: ...57 10 6 2 Sonic Duo Fit ...
Страница 58: ...58 10 6 3 Sonic ...
Страница 59: ...59 10 6 4 Sonic Fit ...
Страница 65: ...65 ...
Страница 66: ...66 ...
Страница 67: ...67 ...
Страница 68: ...68 ...