41
Do not spill liquid disinfectant on the equipment.
Do not use organic solvents, for example, thinner, to clean the equipment.
In the event that the developer solution is spilled on the panel, clean it
immediately, as these solutions may compromise the equipment's paint.
Sterilization parameters must always be followed.
Accessories that are not properly sterilized can cause disease in patients.
All accessories suitable for sterilization must be sterilized only in an autoclave
at 135 ° C with at least 3 minutes of waiting time and pressure of 2.2 bar.
If these items are autoclaved, disinfection by alternative methods is not necessary.
There is no limit on cycles or application time that the equipment and its parts can tolerate during
the cleaning, disinfection and / or sterilization process, following the instructions in this manual.
5.2.BICARBONATE CONTAINER
Locate the bicarbonate container through the side access, remove it by turning it counterclockwise
and clean it with a dry cloth.
Check that the thread is completely free of dust and replace it by turning it clockwise.
Содержание Sonic Duo
Страница 1: ......
Страница 2: ...2 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Страница 24: ...24 Accessories that do not accompany the product 01 07 13 04 10 02 08 14 03 09 15 05 11 06 12 ...
Страница 56: ...56 10 6 EQUIPMENT DIMENSIONS 10 6 1 Sonic Duo ...
Страница 57: ...57 10 6 2 Sonic Duo Fit ...
Страница 58: ...58 10 6 3 Sonic ...
Страница 59: ...59 10 6 4 Sonic Fit ...
Страница 65: ...65 ...
Страница 66: ...66 ...
Страница 67: ...67 ...
Страница 68: ...68 ...