53
10.TECHNICAL SPECIFICATIONS
10.1.EQUIPMENT CLASSIFICATION
10.2.APPLIANCE INFORMATION (GENERAL)
Class of classification according to ANVISA
Class I
Classification of equipment according to standard EN IEC 60601-1
Product classification for applied parts - Type B
Protection Against Electric Shock - Class I
Protection Against Harmful Water Penetration
Dental Prophylaxis by Sodium Bicarbonate / Ultrasound
IP00 - Product not protected against harmful penetration of water and particulate material
Pedal
IP01 - Product not protected against harmful penetration of drops in water and particulate material
Degree of safety of application in the presence of a flammable anesthetic mixture with air,
oxygen or nitrous oxide
Unsuitable equipment
Operation mode
Ultrasound
Non-continuous operation
Time on: 1 min. / Time off: 4 min.
Bicarbonate Jet
Continuous operation
Connection to suitable power supply
External power supply
Power supply voltage
127 / 220 V~ (Dual voltage)
Power supply frequency
50 / 60 Hz
Allowable fluctuation
+/- 10 %
Number of phases
Biphasic
The pedal cannot be installed in an emergency room.