TENSPros
34
Chapter 12: TROUBLESHOOTING
Problem
Possible
Cause
Displays fail to
light up
Make sure that the
batteries are full and
were inserted with the
correct polarity
There is an error in the
electronics
Try new batteries.
Reinsert batteries, ensure
batteries are inserted correctly.
Remove the batteries and
reinsert them after approx.
3 seconds
Stimulation weak
or cannot feel
any stimulation
Electrodes Dried out or
contaminated
Lead wires
Old/worn/damaged
Replace
Replace
Intermittent
output
Lead wires
Verify connection is secure.
Insure firmly
Turn down the intensity.
Rotate lead wires in socket 90°.
If still intermittent, replace lead
wires.
If still intermittent after replacing
lead wires, a component may
have failed. Call the repair
department.
Intermittent
output
Program option in
use
Some programs will seem
intermittent. This is expected.
Option Controls in the Operation
section for a description of the
program option.
Stimulation is
uncomfortable
Intensity is too high
Decrease intensity
Damaged or worn
electrodes or lead wires
Replace