TENSPros
17
Chapter 6:
THERAPY PROGRAM AND WAVEFORMS
TENS PROGRAM
No.
MECHANISM
TECHNICAL DESCRIPTION
01
Program
Five TENS program: B(Burst), N(Normal),
M(Modulation),SD1(Strength Duration 1),
SD2 (Strength Duration 2)
02
Burst program (B)
Burst rate: Adjustable, 0.5 ~ 5Hz
Pulse width adjustable, 50~300μs
Frequency fixed = 100 Hz
03
04
05
Normal program
(N)
The pulse rate and pulse width are adjustable.
It generates continuous stimulation based on
the setting value.
Modulation
program (M)
Modulation mode is a combination of pulse rate
and pulse width modulation. The pulse rate and
width are automatically varied in a cycle pattern.
The pulse width is decreased by 50% from its
original setting in 0.5s, and then the pulse rate
is decreased by 50% from its original setting in
0.5 seconds. Total cycle time is 1 second. In
this mode, pulse rate (2~150Hz) and pulse
width (50~300μs) are fully adjustable.
Strength-Duration
program (SD1)
The SD1(Strength-Duration) mode consists of
automatic modulation intensity and pulse width
in 40% range. The intensity is always increasing
while the pulse width is decreasing and vice-
versa. The intensity is decreased by 40% while
the pulse width is increased by 40% in 5
seconds. In the next 5 seconds, the intensity
is increased by 40% while the pulse width is
decreased by 40%. Total cycle time is 10