TENSPros
35
Electrode active area
size is too small
May not be operating
the device according
to the manual.
Replace electrodes with
large more area
Please check the manual
before use.
Stimulation is
ineffective
Improper electrode and
applicator placement
Unknown
Reposition electrode and
applicator
Contact Clinician
The skin becomes
red and/or you feel
a stabbing pain
The electrodes aren't
stuck onto the skin
properly.
Re-position the electrodes. If
at any time you feel pain or
discomfort stop use immediately.
Ensure the electrode is stuck
securely on the skin.
Replace new electrode.
Output current
stops during
therapy
The electrode pads
come off the skin.
Turn off the device and stick
the electrode pad firmly to
the skin.
Turn off the device and connect
the cable.
Use the electrodes on the
same site every time.
The electrodes are dirty.
The surface of the
electrode was scratched.
Clean the electrode pads with a
damp, lint free cloth or replace
new electrode pads.
The cable is
disconnected
The power of the
batteries has been
exhausted.
Please replace them with
new batteries.
Stimulation is
uncomfortable