TENSPros
21
CH1
CH2
Alternate (A)
No.
MECHANISM
TECHNICAL DESCRIPTION
01
Program
Three IF Modes: N (Normal), M (Modulation)
and M1 (Modulation 1)
02
Normal
program (N)
The pulse rates of the CH1 and CH3 are fixed
in 4000Hz; The interference frequencies of CH2
and CH4 are adjustable from 1Hz to 150Hz.
The pulse rates of CH2 and CH4 are based on
the setting value and fixed frequency (4000Hz),
means that pulse rate=Interference
frequency.
03
Modulation
program (M)
The pulse rates of the CH1 and CH3 is fixed
in 4000Hz; The pulse rates of CH2 and CH4
are based on the interference frequencies and
fixed frequency (4000Hz), means that pulse
rate = Interference frequency. The
interference frequencies are adjustable from
1Hz to 80Hz.
The pulse rates of CH2 and CH4 are modulated
between setting value and 4001Hz in 10 seconds.
IF (Interferential) PROGRAM