Analysis
R&S
®
ZNL
63
User Manual 1178.5989.02 ─ 06
"Linear"
The power level values are converted into linear units prior to averag-
ing. After the averaging, the data is converted back into its original
unit.
"Logarithmic"
For logarithmic scaling, the values are averaged in dBm. For linear
scaling, the behavior is the same as with linear averaging.
"Power"
Activates linear power averaging.
The power level values are converted into unit Watt prior to averag-
ing. After the averaging, the data is converted back into its original
unit.
Use this mode to average power values in Volts or Amperes correctly.
In particular, for small VBW values (smaller than the RBW), use
power averaging mode for correct power measurements in FFT
sweep mode.
Remote command:
Predefined Trace Settings - Quick Config
Commonly required trace settings have been predefined and can be applied very
quickly by selecting the appropriate button.
Function
Trace Settings
Preset All Traces
Trace 1:
Clear Write
Traces 2-6:
Blank
Set Trace Mode
Max | Avg | Min
Trace 1:
Max Hold
Trace 2:
Average
Trace 3:
Min Hold
Traces 4-6:
Blank
Set Trace Mode
Max | ClrWrite | Min
Trace 1:
Max Hold
Trace 2:
Clear Write
Trace 3:
Min Hold
Traces 4-6:
Blank
Trace 1 / Trace 2 / Trace 3 / Trace 4 (Softkeys)
Displays the "Traces" settings and focuses the "Mode" list for the selected trace.
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>[:STATe]
Copy Trace
Access
: "Overview" > "Analysis" > "Traces" > "Copy Trace"
Or: [TRACE] > "Copy Trace"
Copies trace data to another trace.
Trace Settings