The "Sensor" connector is provided by the "Additional Interfaces" option R&S
FPL1-
B5. Additionally, the power sensor measurement requires the option R&S
FPL1-K9.
....................................................................................... 35
............................................................................................35
How to Work With a Power Sensor
.........................................................................39
5.3.3.1
Basics on Power Sensors
For precise power measurement, up to 4 power sensors can be connected to the
instrument via the optional power sensor interface (on the rear panel) or the USB con-
nectors. Both manual operation and remote control are supported.
For a detailed list of supported sensors, see the data sheet.
Signal
source
Power
sensor
R&S
®
FPL
Figure 5-2: Power sensor support – standard test setup
Using the power sensor with several applications
The power sensor cannot be used from the R&S
ZNL firmware and the R&S Power
Viewer Plus (virtual power meter for displaying results of the R&S NRP power sensors)
simultaneously.
Result display
The results of the power sensor measurements are displayed in the marker table. For
each power sensor, a row is inserted. The sensor index is indicated in the "Type" col-
umn.
5.3.3.2
Power Sensor Settings
Access
: "Overview" > "Input" > "Power Sensor" tab
The power sensor measurement requires the option R&S
FPL1-K9.
Each sensor is configured on a separate tab.
Receiving Data Input and Providing Data Output