Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
133
User Manual 1178.5989.02 ─ 06
TRACe:IQ:EGATe:GAP
<Samples>
This command defines the interval between several gate periods for gated measure-
ments with the I/Q analyzer.
Parameters:
<Samples>
<numeric value>
Max = (440 MS * sample rate/200MHz) -1
pretrigger samples defined by
sample rate defined by
Range:
1…Max (samples)
*RST:
1
Example:
TRAC:IQ:EGAT:GAP 2
TRACe:IQ:EGATe:LENGth
<GateLength>
This command defines the gate length for gated measurements with the I/Q analyzer.
Parameters:
<GateLength>
<numeric value>
Max = (440 MS * sample rate/200MHz) -1
pretrigger samples defined by
sample rate defined by
Range:
1…Max (samples)
*RST:
100
Example:
TRAC:IQ:EGAT:LENG 2000
TRACe:IQ:EGATe:NOF
<Number>
This command defines the number of gate periods after the trigger signal for gated
measurements with the I/Q analyzer.
Parameters:
<Number>
Range:
1 to 1023
*RST:
1
Example:
TRAC:IQ:EGAT:NOF 2
TRACe:IQ:EGATe:TYPE
<Type>
This command selects the gate mode for gated measurements with the I/Q analyzer.
Note
: The IF power trigger holdoff time is ignored if you are using the "Level" gate
mode in combination with an IF Power trigger.
Parameters:
<Type>
LEVel
EDGE
*RST:
EDGE
Configuring I/Q Analyzer Measurements