Configuration
R&S
®
ZNL
59
User Manual 1178.5989.02 ─ 06
To determine the required reference level, a level measurement is performed on the
R&S
ZNL.
If necessary, you can optimize the reference level further. Decrease the attenuation
level manually to the lowest possible value before an overload occurs, then decrease
the reference level in the same way.
You can change the measurement time for the level measurement if necessary (see
Changing the Automatic Measurement Time ( MeastimeManual )"
Remote command:
Resetting the Automatic Measurement Time ( MeastimeAuto )
Resets the measurement duration for automatic settings to the default value.
Remote command:
[SENSe:]ADJust:CONFigure:DURation:MODE
Changing the Automatic Measurement Time ( MeastimeManual )
This function allows you to change the measurement duration for automatic setting
adjustments. Enter the value in seconds.
Note:
The maximum possible measurement duration depends on the currently
selected measurement and the installed (optional) hardware. Thus, the measurement
duration actually used to determine the automatic settings may be shorter than the
value you define here.
Remote command:
[SENSe:]ADJust:CONFigure:DURation:MODE
[SENSe:]ADJust:CONFigure:DURation
Upper Level Hysteresis
When the reference level is adjusted automatically using the
internal attenuators are also adjusted. To avoid frequent adaptation due to small
changes in the input signal, you can define a hysteresis. This setting defines an upper
threshold the signal must exceed (compared to the last measurement) before the refer-
ence level is adapted automatically.
Remote command:
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer
Lower Level Hysteresis
When the reference level is adjusted automatically using the
internal attenuators are also adjusted. To avoid frequent adaptation due to small
changes in the input signal, you can define a hysteresis. This setting defines a lower
threshold the signal must fall below (compared to the last measurement) before the ref-
erence level is adapted automatically.
Remote command:
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer
Adjusting Settings Automatically