Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
205
User Manual 1178.5989.02 ─ 06
................................................................................205
..................................................................................... 207
................................................................................... 210
.....................................................................................212
9.8.1
Retrieving Captured I/Q Data
The raw captured I/Q data is output in the form of a list.
............................................................................................... 205
............................................................................................. 205
TRACe:IQ:DATA
Return values:
<Results>
TRACe:IQ:DATA:FORMat
<Format>
This command selects the order of the I/Q data.
For details see
Chapter B, "Reference: Format Description for I/Q Data Files"
Parameters:
<Format>
COMPatible | IQBLock | IQPair
COMPatible
I and Q values are separated and collected in blocks: A block
(512k) of I values is followed by a block (512k) of Q values, fol-
lowed by a block of I values, followed by a block of Q values etc.
(I,I,I,I,Q,Q,Q,Q,I,I,I,I,Q,Q,Q,Q...)
IQBLock
First all I-values are listed, then the Q-values
(I,I,I,I,I,I,...Q,Q,Q,Q,Q,Q)
IQPair
One pair of I/Q values after the other is listed
(I,Q,I,Q,I,Q...).
*RST:
IQBL
TRACe:IQ:DATA:MEMory?
[<OffsetSamples>,<NoOfSamples>]
This command queries the I/Q data currently stored in the capture buffer of the
R&S
ZNL.
By default, the command returns all I/Q data in the memory. You can, however, narrow
down the amount of data that the command returns using the optional parameters.
Retrieving Results