
Configuration
R&S
®
ZNL
37
User Manual 1178.5989.02 ─ 06
The detected
serial numbers
of the power sensors connected to the instrument are
provided in a selection list. For each of the four available power sensor indexes
( "Power Sensor 1" ... "Power Sensor 4" ), which correspond to the tabs in the configu-
ration dialog, one of the detected serial numbers can be assigned. The physical sensor
is thus assigned to the configuration setting for the selected power sensor index.
By default, serial numbers not yet assigned are automatically assigned to the next free
power sensor index for which "Auto Assignment" is selected.
Alternatively, you can assign the sensors manually by deactivating the "Auto" option
and selecting a serial number from the list.
Remote command:
SYSTem:COMMunicate:RDEVice:PMETer<p>:DEFine
SYSTem:COMMunicate:RDEVice:PMETer<p>:CONFigure:AUTO[:STATe]
SYSTem:COMMunicate:RDEVice:PMETer<p>:COUNt?
Zeroing Power Sensor
Starts zeroing of the power sensor.
For details on the zeroing process refer to the R&S
ZNL User Manual.
Remote command:
CALibration:PMETer<p>:ZERO:AUTO ONCE
Frequency Manual
Defines the frequency of the signal to be measured. The power sensor has a memory
with frequency-dependent correction factors. This allows extreme accuracy for signals
of a known frequency.
Remote command:
Frequency Coupling
Selects the coupling option. The frequency can be coupled automatically to the center
frequency of the instrument or to the frequency of marker 1.
Remote command:
[SENSe:]PMETer<p>:FREQuency:LINK
Unit/Scale
Selects the unit with which the measured power is to be displayed. Available units are
dBm, dB, W and %.
If dB or % is selected, the display is relative to the reference value that is defined with
either the "Meas -> Ref" setting or the "Reference Value" setting.
Remote command:
Receiving Data Input and Providing Data Output