Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
206
User Manual 1178.5989.02 ─ 06
If no parameters are specified with the command, the entire trace data is retrieved; in
this case, the command returns the same results as
that the
TRAC:IQ:DATA?
command initiates a new measurement before returning the
captured values, rather than returning the existing data in the memory.)
The command returns a comma-separated list of the measured values in floating point
format (comma-separated values = CSV). The number of values returned is 2 * the
number of complex samples.
The total number of complex samples is displayed in the channel bar in manual opera-
tion and can be calculated as:
<SampleRate> * <CaptureTime>
(See
Query parameters:
<OffsetSamples>
Selects an offset at which the output of data should start in rela-
tion to the first data. If omitted, all captured samples are output,
starting with the first sample.
Range:
0 to <# of samples> – 1, with <# of samples> being
the maximum number of captured values
*RST:
0
<NoOfSamples>
Number of samples you want to query, beginning at the offset
you have defined. If omitted, all captured samples (starting at
offset) are output.
Range:
1 to <# of samples> - <offset samples> with <# of
samples> maximum number of captured values
*RST:
<# of samples>
Return values:
<IQData>
Measured value pair (I,Q) for each sample that has been recor-
ded.
By default, the first half of the list contains the I values, the sec-
ond half the Q values. The order can be configured using
The data format of the individual values depends on
Default unit: V
Retrieving Results