Function Test
R&S FSMR
1313.9617.12
1.20
E-2
Power-Save Mode
Display:
The R&S FSMR offers the possibility of switching on a power-save mode for the screen display. The
backlighting will be switched off if no entry is made on the front panel (key, softkey or hotkey as well as
spinwheel) during the selected response time.
In order to switch on the power-save mode:
1. Call the
DISPLAY - CONFIG DISPLAY
submenu to configure the screen display:
Press
DISP
key
Press
CONFIG DISPLAY
softkey
2. Activate the save mode
Press
DISPLAY PWR SAVE
softkey.
The softkey is highlighted in colour, thus indicating that the power-save mode is on. At the
same time the data entry for the delay time is opened.
3. Define the delay time
Enter the required response time in minutes and confirm the entry using the
ENTER
key.
The screen will be blanked out after the selected time period has elapsed.
Hard disk:
A power-save mode is preset for the built-in hard disk which is automatically closed down 15 minutes
after the last access.
Recalling the Most Recent Instrument Settings
The R&S FSMR stores its current instrument settings onto the hard disk every time it is switched off via
the ON/STANDBY key. After each power-on, the R&S FSMR is reloaded with the operational
parameters which were active just prior to the last power-off (STANDBY or AC power OFF) or were set
with STARTUP RECALL (see Chapter 4 "Saving and Recalling Data Sets").
Storing the current instrument settings is not possible if the instrument is switched off
using the POWER ON switch at the rear panel or when unplugging the mains cord.
After power-on the instrument settings stored previously on the hard disk will be
loaded in this case.
Function Test
After turning on the AC power, the R&S FSMR will display the following message on the display screen:
Rohde & Schwarz GmbH & Co. KG
Analyzer BIOS
Vx.y
After appearance of the above message, a selftest of the controller hardware is performed.
Subsequently, the Windows XP controller boots and the measurement screen will appear.
The system self-alignment is activated via
CAL
key,
CAL TOTAL
softkey. The individual results of the
self-alignment (PASSED / FAILED) can be displayed in the
CAL
menu (
CAL RESULTS)
.
With the aid of the built-in selftest functions (
SETUP
key,
SERVICE, SELFTEST
soft keys), the
functional integrity of the instrument can be verified and/or defective modules can be localized.
Содержание FSMR series
Страница 1: ...R S FSMR Measuring Receiver Getting Started Test Measurement Getting Started 1313 9698 02 03 ...
Страница 7: ......
Страница 9: ......
Страница 11: ...Contents Preparing for Operation R S FSMR 1313 9723 12 I 1 2 E 1 Fig 1 1 Front View ...
Страница 13: ...Front View R S FSMR 1313 9723 12 1 2 E 2 Fig 1 1 Front View ...
Страница 15: ...Front View R S FSMR 1313 9723 12 1 4 E 2 Fig 1 1 Front View ...
Страница 17: ...Front View R S FSMR 1313 9723 12 1 6 E 2 Fig 1 1 Front View ...
Страница 19: ...Front View R S FSMR 1313 9723 12 1 8 E 2 Fig 1 1 Front View ...
Страница 51: ......
Страница 71: ......
Страница 102: ...R S FSMR Menu Overview 1166 3363 12 3 29 E 3 AMPT Key ...
Страница 107: ...Menu Overview R S FSMR 1166 3363 12 3 34 E 3 DEMOD Key ...
Страница 108: ...R S FSMR Menu Overview 1166 3363 12 3 35 E 3 AUDIO Key ...
Страница 111: ...Menu Overview R S FSMR 1166 3363 12 3 38 E 3 AMPT Key ...
Страница 112: ...R S FSMR Menu Overview 1166 3363 12 3 39 E 3 MEAS Key ...
Страница 113: ...Menu Overview R S FSMR 1166 3363 12 3 40 E 3 ...
Страница 114: ...R S FSMR Menu Overview 1166 3363 12 3 41 E 3 BW Key ...
Страница 116: ...R S FSMR Menu Overview 1166 3363 12 3 43 E 3 Menu Overview another Keys MKR Key ...
Страница 120: ...R S FSMR Menu Overview 1166 3363 12 3 47 E 3 TRACE Key ...
Страница 123: ...Menu Overview R S FSMR 1166 3363 12 3 50 E 3 FILE Key ...
Страница 125: ...Menu Overview R S FSMR 1166 3363 12 3 52 E 3 SETUP Key ...
Страница 127: ...Menu Overview R S FSMR 1166 3363 12 3 54 E 3 LOCAL Menu LOCAL ...
Страница 128: ...R S FSMR Menu Overview 1166 3363 12 3 55 E 3 Menu Overview Ext Generator Control ...