List of Commands/Annex C
CTS30
1094.0006.30 C.18
E-4
1.5.1.2 BER Measurement
Measured values of class Ib
Syntax:
Execute new
measurement and signal
result
Read result only
READ[:SCALar]:BER:CLIB[:BER]?
READ[:SCALar]:BER:CLIB:RBER?
FETCh[:SCALar]:BER:CLIB[:BER]?
FETCh[:SCALar]:BER:CLIB:RBER?
Return:
BER and RBER:
0 ... 100
(Unit: %)
State:
MCE
Option:
none Query only
Note:
No default values, (N4)
If a read is performed on BER, all subsequent fetches will return BER values
until a RBER read is carried out.
The same applies for RBER
Measured values of class II
Syntax:
Execute new
measurement and signal
result
Read result only
READ[:SCALar]:BER:CLII[:BER]?
READ[:SCALar]:BER:CLII:RBER?
FETCh[:SCALar]:BER:CLII[:BER]?
FETCh[:SCALar]:BER:CLII:RBER?
Return:
BER and RBER:
0 ... 100
(Unit: %)
State:
MCE
Option:
none Query only
Note:
No default values, (N4)
If a read is performed on BER, all subsequent fetches will return BER values
until a RBER read is carried out.
The same applies for RBER
Measured values of erased frames
Syntax:
Execute new
measurement and signal
result
Read result only
READ[:SCALar]:BER:EFRames[:FER]?
FETCh[:SCALar]:BER:EFRames[:FER]?
Return:
BER and RBER:
0 ... 100
(Unit: %)
State:
MCE
Option:
none Query only
Note:
No default values, (N4)