Command Description
R&S FSH-K1
1157.3564.12 52
E-8
Channel Power
Command
Parameters
Unit
Comment
CHPWR
<numeric value>
GET command only
CHPWRSTD 0…3
CHPWRCSTD <string>
CHPWRUNIT 0…2
CHPWRBW <numeric
value>
Hz
LVLADJUST
CMD command only
CHPWR (Channel Power)
This command queries the measured channel power.
Example:
get<cr>
0<cr>
chpwr<cr>
‘query
power
level
from
sensor
0<cr>
‘responds parameter ok
-47.45<cr>
‘response: channel power
CHPWRSTD (Channel Power Standard)
This command defines the standard used for the channel power measurement.
Value
Channel Power Standard
0
User
1
3GPP WCDMA
2
cdmaOne
3
cdma2000 1x
Example:
set<cr>
0<cr>
chpwrstd,2<cr>
‘select cdmaOne as standard
0<cr>
‘responds parameter ok
Note: If any customized standard was previously selected a value of ‘4’ will be returned as a response
to a GET CHPWRSTD query.
CHPWRCSTD (Channel Power Customized Standard)
This command selects the customized standard for the channel power measurement previously loaded
with R&S FSHView software.
Example:
set<cr>
0<cr>
chpwrcstd,MyStd<cr>
‘select MyStd as standard
0<cr>
‘responds parameter ok
Содержание 1157.3458.02
Страница 4: ......
Страница 5: ...R S FSH K1 Table of Contents 1157 3564 12 I 1 E 8...
Страница 6: ......