R&S FSH-K1
Command Description
1157.3564.12 43
E-8
LIMDEF (Define Limit Line)
This command defines a limit line.The list of parameters are the following:
<name>,<description>,<x-unit>,<x-scale>,<y-unit>,<x0>,<y0>[,...<xn>,<yn>]
The parameter <name> and <description> are strings. The parameters x-unit and y-unit are listed in the
tables below:
Parameter x-unit:
Value
Unit
0
Hz
1
Seconds
2
Meter
The parameter x-scale defines whether the x values are absolute values or relative values according to
the center x value:
Value
x-scale
0
Absolute
1
Relative
Parameter y-unit:
Value
Unit
0
dB
1
dBm
2
dBµV
3
dBmV
4
dBµV/m
5
dBµA/m
6
VSWR
7
Rho
8
Volt
9
Watt
10 Volt/m
11 Watt/m
2
12 Seconds
13 Degrees
The following example defines the limit line “LIMIT”, with description “Max” in dBm on the frequency
axis (Hz) as an absolute limit line with 4 values (100MHz -30dBm, 200MHz -10dBm, 300MHz -10dBm,
400MHz -30dBm):
Example:
set<cr>
0<cr>
limdef,LIMIT,Max,0,0,1,100e6,-30,200e6,-10,300e6,
-10,400e6,-30<cr>
‘defines limit line “LIMIT”
0<cr>
‘responds parameter ok
Note: To replace an existing limit line in the instrument it has to be deleted first by using the LIMDEL
command described below.