Command Description
R&S FSH-K1
1157.3564.12 46
E-8
Tracking Generator
This command set applies to model 13, model 23 and model 26 only.
Command
Parameters
Unit
Comment
CAL_TGSCLRFL
CMD command only
CAL_TGSCLTRN
CMD command only
CAL_TGVECRFL
CMD command only
CAL_TGVECTRN
CMD command only
TRANSCAL
GET command only
REFLCAL
GET command only
TRANSVECTCAL
GET command only
Option R&S FSH-K2
REFLVECTCAL
GET command only
Option R&S FSH-K2
TGATT
<numeric value>
Model 23 & Model 26
(Serial No 100500 or higher)
TGLVL
<numeric value>
Model 23 only
TGMODE
[0…4]
Option R&S FSH-K2
CABLELOSS
GET command only
Option R&S FSH-K2
ELCABLENVAL
<numeric value>
m / feet
GET command only
Option R&S FSH-K2
CAL_TGSCLRFL (Calibrate Tracking Generator Scalar Reflection)
This command initiates a scalar reflection calibration. This calibration contains two phases; OPEN
phase and SHORT phase. Connect OPEN before initiating calibration command, after first phase
reports ready connect SHORT and continue by issuing the command again.
Example:
cmd<cr>
0<cr>
cal_tgsclrfl<cr>
‘initiate scalar reflection calibration (OPEN)
0<cr>
‘responds parameter ok
0<cr>
‘first phase ready
cal_tgsclrfl<cr>
‘continue scalar reflection calibration (SHORT)
0<cr>
‘responds parameter ok
0<cr>
‘scalar reflection calibration ready
CAL_TGSCLTRN (Calibrate Tracking Generator Scalar Transmission)
This command initiates a scalar transmission calibration. This calibration contains one phase;
calibrating THROUGH. Connect THROUGH before initiating calibration command.
Example:
cmd<cr>
0<cr>
cal_tgscltrn<cr>
‘initiate scalar transmission calibration
0<cr>
‘responds parameter ok
0<cr>
‘scalar transmission calibration ready