R&S FSH-K1
Command Description
1157.3564.12 15
E-8
EXTREF (External Reference status)
This command returns the external reference status.
Value
Status external reference
0
Disabled
1
Out of range
2
Catching
3
Locked
Example:
get<cr>
0<cr>
extref<cr>
‘get status of external reference
0<cr>
‘responds parameter ok
1<cr>
‘response status out of range
DISPLAY
This command turns the display on or off.
Value
0
Off
1
On
Example:
set<cr>
0<cr>
display,0<cr>
‘turn
display
off
0<cr>
‘responds parameter ok
TEMP (Temperature)
This command returns the current temperature inside the instrument.
Example:
get<cr>
0<cr>
temp<cr>
‘query
temperature
0<cr>
‘responds parameter ok
32.6
‘instrument temperature (Celsius)