Operator’s Manual · Version 3.1
9 Messages and alarms
Cedex Bio System
List of alarm messages
7066
An internal control unit (%d) reported a wrong firmware
version.
Reinstall the system software.
7067
An internal control unit (%d) reported a checksum or an
attach error.
Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
7068
[%d] Wash aborted due to open cover.
Close the cover and restart the system.
7069
[%d] Transfer initialization failed due to an open main
cover.
Ensure that all covers are closed. Restart the system.
7070
The available reagent bottle volume in bottle set %s is not
sufficient. The bottle set has been marked as empty.
Replace the bottle set with a new one.
7071
Photometer error. Signal is invalid or unstable.
Switch off the instrument. Remove and clean the rotor. If
the error recurs perform <Replace Photometer Lamp>
and <Abs Air/Water Calibration>. If the error persists call
service.
7072
[%d] The reagent mixing failed due to a level detection
problem.
Check if the probe is obstructed. Check the filling level of
the associated reagent bottle. If the error persists call ser-
vice.
7073
[%s] Peristaltic pump speed is outside the range.
Check ISE tubing for obstructions and leaks. Ensure that
the electrode tension lever and the peristaltic pump cover
are closed. Perform <Initialize ISE Unit>. If the error per-
sists call service.
7074
[%s] The detected sample or calibrator segment is too
short.
Check the ISE tubing for obstructions and leaks. Ensure
that the electrodes are installed correctly. If the error per-
sists call service.
7075
[%s] A timeout error occurred while performing an ISE
measurement.
Restart the system and repeat the measurement. If the
error persists call service.
7076
[%s] Initialization of the ISE unit failed due to a fluid
transport problem.
Check the ISE tower and the ISE tubing for obstructions
and leaks. Check the electrodes. Perform <Initialize ISE
Unit>. If the error persists call service.
7083
[%s] The ISE bypass tubing is blocked.
Perform <Replace ISE Unit Tubing>. Follow the instruc-
tions in the Operator's Manual. If the error persists call
service.
7084
[%s] An ISE pinch valve is probably defective.
Restart the system. If the error persists call service.
7086
[%s] The ISE air pressure is outside the range.
Check the air tubing underneath the ISE tower for
obstructions. If the error persists call service.
7087
[%s] ISE air pressure sensor detected that the pressure is
too low.
Restart the system. If the problem persists, call service.
7088
[%s] A failure of the internal ISE power supply has been
detected.
Restart the system. If the error persists call service.
7099
[%s] Weight sensors of the ISE Ref and ISE Cal bottle are
not calibrated correctly.
In <Diagnostics> perform <Adjust ISE Ref Bottle Sensor>
and <Adjust ISE Cal Bottle Sensor>. If the error persists
call service.
7100
A power fail has been detected and an automatic system
restart was performed.
7101
[%s] Rotor position error. Switch off the instrument.
Check if the rotor is blocked or jamming. Eliminate the
cause. Restart the system. If the error persists call service.
7102
[%s] Transfer bumper init failed. Perform restart.
ID
Message
Comment
Table D-1
List of alarm messages (Continued)
Содержание Cedex Bio
Страница 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Страница 10: ......
Страница 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Страница 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Страница 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Страница 132: ......
Страница 303: ...8 General maintenance C 3 Maintenance C...
Страница 304: ......
Страница 306: ...Roche Diagnostics C 4 Operator s Manual Version 3 1 8 General maintenance Cedex Bio System Table of contents...
Страница 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Страница 340: ......
Страница 342: ...Roche Diagnostics D 4 Operator s Manual Version 3 1 9 Messages and alarms Cedex Bio System Table of contents...
Страница 378: ...Roche Diagnostics D 40 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Table of contents...
Страница 390: ...Roche Diagnostics D 52 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Detailed procedures...
Страница 391: ...12 ISE description E 3 13 ISE operation E 15 14 ISE maintenance E 39 15 ISE troubleshooting E 67 ISE E...
Страница 392: ......
Страница 394: ...Roche Diagnostics E 4 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Table of contents...
Страница 404: ...Roche Diagnostics E 14 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Technical specifications...
Страница 406: ...Roche Diagnostics E 16 Operator s Manual Version 3 1 13 ISE operation Cedex Bio System Table of contents...
Страница 430: ...Roche Diagnostics E 40 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System Table of contents...
Страница 456: ...Roche Diagnostics E 66 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System ISE maintenance actions...
Страница 458: ...Roche Diagnostics E 68 Operator s Manual Version 3 1 15 ISE troubleshooting Cedex Bio System Table of contents...
Страница 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Страница 472: ......
Страница 480: ...Roche Diagnostics F 10 Operator s Manual Version 3 1 Glossary Cedex Bio System worklist worklist...
Страница 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Страница 493: ...Revisions G...
Страница 494: ......
Страница 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Страница 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...