Roche Diagnostics
Operator’s Manual · Version 3.1
D-11
Cedex Bio System
9 Messages and alarms
List of alarm messages
0020
This reagent set has more bottles than there is space
available on the disk.
Load another disk or remove unused reagents.
0021
Disk space problem.
0022
You are about to delete results that have been printed or
sent to host!
0023
Enter the A1c part.
0024
Now enter the A1c part.
5001
Air detected in the fluid system. Water container or
cleaner bottle is empty.
5002
[%d] Air detected in the fluid system.
Refill the water container. Check the filling level of the
cleaner bottle. Replace the cleaner bottle if required. Per-
form <Prime Fluid System>. If the error persists call ser-
vice.
5003
[%s] An internal software error occurred. A parameter is
not valid.
Restart the system. If the error persists call service.
5004
[%s] Error syringe motor. The syringe motor could not be
started.
Restart the system. If the error persists call service.
5010
[%s] Initialization of syringe failed.
Switch off instrument. Remove right side panel. Check for
syringe mechanism jamming or blockage. Ensure correct
syringe installation. Restart the system. If the error per-
sists call service.
5013
[%s] Error mixer motor. The mixer motor could not be
started.
Restart the system. If the error persists call service.
5019
[%s] Initialization of the mixer motor failed.
Switch off the instrument. Remove the transfer head
cover. Check if the mixing mechanism is jamming or
blocked. Restart the system. If the error persists call ser-
vice.
5022
[%s] Error temperature sensor analyzer heating system.
The sensor does not react anymore.
Restart the system. If the error persists call service.
5024
[%s] Error temperature sensor analyzer cooling system.
The sensor does not react anymore.
Restart the system. If the error persists call service.
5026
[%s] Error temperature sensor cooling system. The heat
sink temperature sensor does not react anymore.
Restart the system. If the error persists call service.
5028
[%s] Error temperature sensor instrument. The ambient
temperature sensor does not react anymore.
Restart the system. If the error persists call service.
5030
[%s] Error wash pump P1. The wash pump could not be
started.
Restart the system. If the error persists call service.
5032
[%s] Error waste pump P2. The waste pump could not be
started.
Restart the system. If the error persists call service.
5034
[%s] Error wash valve V1. The valve does not react
anymore.
Restart the system. If the error persists call service.
5036
[%s] Error cleaner valve V2. The valve does not react
anymore.
Restart the system. If the error persists call service.
5038
[%s] Error analyzer heating system. The heating system
does not react anymore.
Restart the system. If the error persists call service.
5040
[%s] Error cooling system. One of the Peltier elements
does not react anymore.
Restart the system. If the error persists call service.
ID
Message
Comment
Table D-1
List of alarm messages (Continued)
Содержание Cedex Bio
Страница 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Страница 10: ......
Страница 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Страница 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Страница 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Страница 132: ......
Страница 303: ...8 General maintenance C 3 Maintenance C...
Страница 304: ......
Страница 306: ...Roche Diagnostics C 4 Operator s Manual Version 3 1 8 General maintenance Cedex Bio System Table of contents...
Страница 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Страница 340: ......
Страница 342: ...Roche Diagnostics D 4 Operator s Manual Version 3 1 9 Messages and alarms Cedex Bio System Table of contents...
Страница 378: ...Roche Diagnostics D 40 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Table of contents...
Страница 390: ...Roche Diagnostics D 52 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Detailed procedures...
Страница 391: ...12 ISE description E 3 13 ISE operation E 15 14 ISE maintenance E 39 15 ISE troubleshooting E 67 ISE E...
Страница 392: ......
Страница 394: ...Roche Diagnostics E 4 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Table of contents...
Страница 404: ...Roche Diagnostics E 14 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Technical specifications...
Страница 406: ...Roche Diagnostics E 16 Operator s Manual Version 3 1 13 ISE operation Cedex Bio System Table of contents...
Страница 430: ...Roche Diagnostics E 40 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System Table of contents...
Страница 456: ...Roche Diagnostics E 66 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System ISE maintenance actions...
Страница 458: ...Roche Diagnostics E 68 Operator s Manual Version 3 1 15 ISE troubleshooting Cedex Bio System Table of contents...
Страница 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Страница 472: ......
Страница 480: ...Roche Diagnostics F 10 Operator s Manual Version 3 1 Glossary Cedex Bio System worklist worklist...
Страница 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Страница 493: ...Revisions G...
Страница 494: ......
Страница 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Страница 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...