Roche Diagnostics
Operator’s Manual · Version 3.1
D-19
Cedex Bio System
9 Messages and alarms
List of alarm messages
7141
[%s] Cl calibrator measurement unstable.
Perform <Electrode Service> and <Prime ISE Reference
and Calib.>. Check ISE Cal and Ref flow. Check the ISE
tubing for obstructions. If the error persists call service.
7142
[%s] K calibrator measurement unstable.
Perform <Electrode Service> and <Prime ISE Reference
and Calib.>. Check ISE Cal and Ref flow. Check the ISE
tubing for obstructions. If the error persists call service.
7143
[%d] An internal software problem (%d) occurred. The
access to the internal EEPROM failed.
Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
7144
[%d] The instrument battery is low or not connected. The
date and time may no longer be valid. A replacement of
the battery is probably required.
Call service.
7145
[%d] The action <Clean ISE> has been stopped due to an
open cover.
Ensure that all covers are closed while the system is being
initialized, or in the Operating or Maintenance status.
7146
[%d] The initialization of the system has been stopped due
to an open cover.
Ensure that all covers are closed while the system is being
initialized, or in the Operating or Maintenance status.
7147
[%d] The initialization of the ISE Unit has been stopped
due to an open cover.
Ensure that all covers are closed while the system is being
initialized, or in the Operating or Maintenance status.
7148
[%d] Initialization of the ISE Unit has been stopped due to
a blocked transfer.
Switch off the instrument. Check if the transfer is blocked
or jamming. Ensure that all covers are closed. Restart the
system. If the error persists call service.
7149
[%d] Initialization of the rotor failed. The calculated offset
is outside the range.
Switch off the instrument. Check if the rotor is blocked or
jamming. Restart the system. If the error persists call ser-
vice.
7150
[%d] Initialization of the transfer failed due a level
detection problem.
Switch off the system. Clean the probe and probe holder
and ensure that they are installed correctly. Restart the
system. If the error persists call service.
7151
[%d] Dispensing fluid into the ISE tower failed due to an
incorrect X-transfer or probe position.
Switch off the instrument. Check if the X- transfer and Z-
transfer are blocked or jamming. Restart the system. If the
error persists call service.
7152
[%d] A pipetting or dispensing cycle failed due to an
incorrect X-transfer or probe position.
Switch off the instrument. Check if the X- transfer and Z-
transfer are blocked or jamming. Restart the system. If the
error persists call service.
7153
[%d] A reagent mixing cycle failed because the probe
crashed.
Switch off the instrument. Check if the probe and probe
holder are correctly installed. Replace a bent probe.
Restart the system. If the error persists call service.
7154
[%d] A probe wash cycle failed because the probe crashed. Switch off the instrument. Check if the probe and probe
holder are correctly installed. Replace a bent probe.
Restart the system. If the error persists call service.
7155
[%d] Initialization of the transfer failed because the probe
crashed.
Switch off the instrument. Check if the probe and probe
holder are correctly installed. Replace a bent probe.
Restart the system. If the error persists call service.
7156
[%d] A prime cycle failed because the probe crashed.
Switch off the instrument. Check if the probe and probe
holder are correctly installed. Replace a bent probe.
Restart the system. If the error persists call service.
7157
[%d] Initialization of the Z-transfer failed due to a level
detection problem.
Switch off the instrument. Check if the probe and probe
holder are correctly installed. Clean the init plate. Restart
the system. If the error persists call service.
ID
Message
Comment
Table D-1
List of alarm messages (Continued)
Содержание Cedex Bio
Страница 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Страница 10: ......
Страница 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Страница 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Страница 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Страница 132: ......
Страница 303: ...8 General maintenance C 3 Maintenance C...
Страница 304: ......
Страница 306: ...Roche Diagnostics C 4 Operator s Manual Version 3 1 8 General maintenance Cedex Bio System Table of contents...
Страница 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Страница 340: ......
Страница 342: ...Roche Diagnostics D 4 Operator s Manual Version 3 1 9 Messages and alarms Cedex Bio System Table of contents...
Страница 378: ...Roche Diagnostics D 40 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Table of contents...
Страница 390: ...Roche Diagnostics D 52 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Detailed procedures...
Страница 391: ...12 ISE description E 3 13 ISE operation E 15 14 ISE maintenance E 39 15 ISE troubleshooting E 67 ISE E...
Страница 392: ......
Страница 394: ...Roche Diagnostics E 4 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Table of contents...
Страница 404: ...Roche Diagnostics E 14 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Technical specifications...
Страница 406: ...Roche Diagnostics E 16 Operator s Manual Version 3 1 13 ISE operation Cedex Bio System Table of contents...
Страница 430: ...Roche Diagnostics E 40 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System Table of contents...
Страница 456: ...Roche Diagnostics E 66 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System ISE maintenance actions...
Страница 458: ...Roche Diagnostics E 68 Operator s Manual Version 3 1 15 ISE troubleshooting Cedex Bio System Table of contents...
Страница 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Страница 472: ......
Страница 480: ...Roche Diagnostics F 10 Operator s Manual Version 3 1 Glossary Cedex Bio System worklist worklist...
Страница 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Страница 493: ...Revisions G...
Страница 494: ......
Страница 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Страница 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...