Define the lot values.
Do one of the following:
Defining calibrator definitions and lots
You need to perform the definitions for all calibrators.
a
To define a calibrator lot
1
Choose
Workplace
>
Lot Data
.
2
Press
Calibrators
.
A screen is displayed that lists all installed calibrator lots.
3
Press .
A screen is displayed, asking you to scan the barcode or to type the data manually.
If…
Do this…
The lot data are
available from a
barcode
1. Scan the barcode.
The screen is displayed again that lists all installed control lots.
The new lot is selected.
You want to type
the data manually
1. Press .
2. Type the name and press
.
3. Type the material code and press
.
You find this code in
Utilities
>
Applications
> select a test >
>
Control
.
4. Type the Lot ID and press
.
5. Type the expiration date. Use the date format as indicated on
the screen.
6. Press
to confirm the definitions.
If a lot with the same ID already exists on the system, a screen is
displayed, asking you whether you want to replace the existing
lot.
Press
to confirm the definitions.
7. Press
*Add new test*
.
8. Press .
A screen is displayed that contains all tests that use this control
and that have not yet been assigned to this lot.
9. Select a test and press
.
10. Type the value for the mean concentration and press
.
11. Type the value for the standard deviation.
12. Press
to confirm the definitions.
The screen for handling control lot data is displayed again.
13. Perform steps 7 through 12 for all tests you want to use.
Содержание Cedex Bio
Страница 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Страница 10: ......
Страница 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Страница 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Страница 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Страница 132: ......
Страница 303: ...8 General maintenance C 3 Maintenance C...
Страница 304: ......
Страница 306: ...Roche Diagnostics C 4 Operator s Manual Version 3 1 8 General maintenance Cedex Bio System Table of contents...
Страница 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Страница 340: ......
Страница 342: ...Roche Diagnostics D 4 Operator s Manual Version 3 1 9 Messages and alarms Cedex Bio System Table of contents...
Страница 378: ...Roche Diagnostics D 40 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Table of contents...
Страница 390: ...Roche Diagnostics D 52 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Detailed procedures...
Страница 391: ...12 ISE description E 3 13 ISE operation E 15 14 ISE maintenance E 39 15 ISE troubleshooting E 67 ISE E...
Страница 392: ......
Страница 394: ...Roche Diagnostics E 4 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Table of contents...
Страница 404: ...Roche Diagnostics E 14 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Technical specifications...
Страница 406: ...Roche Diagnostics E 16 Operator s Manual Version 3 1 13 ISE operation Cedex Bio System Table of contents...
Страница 430: ...Roche Diagnostics E 40 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System Table of contents...
Страница 456: ...Roche Diagnostics E 66 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System ISE maintenance actions...
Страница 458: ...Roche Diagnostics E 68 Operator s Manual Version 3 1 15 ISE troubleshooting Cedex Bio System Table of contents...
Страница 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Страница 472: ......
Страница 480: ...Roche Diagnostics F 10 Operator s Manual Version 3 1 Glossary Cedex Bio System worklist worklist...
Страница 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Страница 493: ...Revisions G...
Страница 494: ......
Страница 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Страница 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...