Roche Diagnostics
Operator’s Manual · Version 3.1
D-31
Cedex Bio System
10 Result flags
List of flags
High Abs
Meaning
Excessive absorbance.
The absorbance value to be used for calculation and checks is > 2.0 Abs.
Message ID
5
Priority
1
Recommended actions
CAL
o
Repeat with fresh calibrator.
CNTL
o
If a single control is flagged, rerun with a fresh control.
o
If more than one control is flagged, perform a calibration.
TEST
o
If a single sample result is flagged, rerun with dilution.
o
If the problem persists, replace the reagent set.
High Act
Meaning
High activity.
o
The absorbance change during measurement is above the accepted limit, i.e. the
reaction has taken off too fast.
Message ID
56
Priority
9
Recommended actions
CAL
1. Check if the correct calibrators are placed on the instrument.
2. Repeat with fresh calibrators.
CNTL
1. Check control. Repeat with fresh control.
TEST
1. Rerun with dilution.
Содержание Cedex Bio
Страница 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Страница 10: ......
Страница 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Страница 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Страница 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Страница 132: ......
Страница 303: ...8 General maintenance C 3 Maintenance C...
Страница 304: ......
Страница 306: ...Roche Diagnostics C 4 Operator s Manual Version 3 1 8 General maintenance Cedex Bio System Table of contents...
Страница 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Страница 340: ......
Страница 342: ...Roche Diagnostics D 4 Operator s Manual Version 3 1 9 Messages and alarms Cedex Bio System Table of contents...
Страница 378: ...Roche Diagnostics D 40 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Table of contents...
Страница 390: ...Roche Diagnostics D 52 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Detailed procedures...
Страница 391: ...12 ISE description E 3 13 ISE operation E 15 14 ISE maintenance E 39 15 ISE troubleshooting E 67 ISE E...
Страница 392: ......
Страница 394: ...Roche Diagnostics E 4 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Table of contents...
Страница 404: ...Roche Diagnostics E 14 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Technical specifications...
Страница 406: ...Roche Diagnostics E 16 Operator s Manual Version 3 1 13 ISE operation Cedex Bio System Table of contents...
Страница 430: ...Roche Diagnostics E 40 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System Table of contents...
Страница 456: ...Roche Diagnostics E 66 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System ISE maintenance actions...
Страница 458: ...Roche Diagnostics E 68 Operator s Manual Version 3 1 15 ISE troubleshooting Cedex Bio System Table of contents...
Страница 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Страница 472: ......
Страница 480: ...Roche Diagnostics F 10 Operator s Manual Version 3 1 Glossary Cedex Bio System worklist worklist...
Страница 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Страница 493: ...Revisions G...
Страница 494: ......
Страница 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Страница 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...