Roche Diagnostics
B-96
Operator’s Manual · Version 3.1
6 Special operations
Cedex Bio System
Deleting QC results
Setting up your calibration schedule
The
Cedex Bio
instrument provides convenient software-supported procedures for
the following major calibration scenarios:
o
Performing all calibrations that are currently due
o
Performing all calibrations that fall due within a configurable forecast period
o
Performing individual calibrations
Performing calibration as part of the Prepare wizard includes performing all
calibrations that are due and that will fall due during the forecast period.
The following table lists the configurable definitions and describes their effect.
Deleting QC results
Accepted QC results are normally deleted from the
QC Status
list as part of the daily
end of shift activities, at the same time, results that were generated before the first day
of the previous month are deleted from the
QC History
list.
This section describes how to proceed if you need to delete QC results outside the
normal daily routine.
e
For deleting QC results at the end of a shift, see
Definition item
Configuration path
Effect
Forecast Hours
Utilities
>
Configuration
>
Calibration
Period of time, starting with the current
date, within which calibration falls due.
Defining a forecast period (value other than
zero) allows to concentrate calibration per-
formance to specific days.
Cal Auto Accept
Utilities
>
Configuration
>
Result
Handling
If on, results that are not flagged are auto-
matically accepted.
Results with flags that are marked in a pre-
defined list of acceptable flags would also be
automatically accepted.
Cal Acceptable
Flags
Utilities
>
Configuration
>
Result
Handling
List of calibration relevant flags. Flags that
are marked are automatically accepted if
Cal Auto Accept
is on.
Table B-10
Calibration definitions and their effect
o
QC results are displayed in the
QC Status
list as soon as they are generated. A copy of
the results is also placed in the
QC History
list. This list keeps the results of the previ-
ous and the current months.
o
Deleting QC results, deletes the associated orders as well.
Содержание Cedex Bio
Страница 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Страница 10: ......
Страница 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Страница 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Страница 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Страница 132: ......
Страница 303: ...8 General maintenance C 3 Maintenance C...
Страница 304: ......
Страница 306: ...Roche Diagnostics C 4 Operator s Manual Version 3 1 8 General maintenance Cedex Bio System Table of contents...
Страница 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Страница 340: ......
Страница 342: ...Roche Diagnostics D 4 Operator s Manual Version 3 1 9 Messages and alarms Cedex Bio System Table of contents...
Страница 378: ...Roche Diagnostics D 40 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Table of contents...
Страница 390: ...Roche Diagnostics D 52 Operator s Manual Version 3 1 11 Troubleshooting Cedex Bio System Detailed procedures...
Страница 391: ...12 ISE description E 3 13 ISE operation E 15 14 ISE maintenance E 39 15 ISE troubleshooting E 67 ISE E...
Страница 392: ......
Страница 394: ...Roche Diagnostics E 4 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Table of contents...
Страница 404: ...Roche Diagnostics E 14 Operator s Manual Version 3 1 12 ISE description Cedex Bio System Technical specifications...
Страница 406: ...Roche Diagnostics E 16 Operator s Manual Version 3 1 13 ISE operation Cedex Bio System Table of contents...
Страница 430: ...Roche Diagnostics E 40 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System Table of contents...
Страница 456: ...Roche Diagnostics E 66 Operator s Manual Version 3 1 14 ISE maintenance Cedex Bio System ISE maintenance actions...
Страница 458: ...Roche Diagnostics E 68 Operator s Manual Version 3 1 15 ISE troubleshooting Cedex Bio System Table of contents...
Страница 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Страница 472: ......
Страница 480: ...Roche Diagnostics F 10 Operator s Manual Version 3 1 Glossary Cedex Bio System worklist worklist...
Страница 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Страница 493: ...Revisions G...
Страница 494: ......
Страница 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Страница 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...