47 • Connecting Instruments to the PC Controller Program
Press and release the POWER ENTER button on the instrument. If you do
not press the POWER ENTER button for 20 seconds, the instrument will
automatically continue its warm up sequence in Normal Mode.
•
When CL RMNDR is set to ON and CL EXPRD is set to NOT USE, if the
instrument is due for calibration, it will show a calibration expired failure
screen and the buzzer and LED arrays will pulse.
The instrument will then connect to the PC Controller program and display
the home screen shown above in step 8. The CHARGE LED will begin to
blink green.
•
When CL RMNDR is set to ON and CL EXPRD is set to CONFIRM, if the
instrument is due for calibration, it will show the following screen and beep
and flash the LED arrays for a few seconds indicating that calibration is
due.
When the beeping and flashing stops, press and release the RESET
SILENCE button on the instrument to continue. The instrument will now
connect to the PC Controller program, the home screen shown in step 8
above will be displayed, and the CHARGE LED will begin to blink green. If
you do not press and release the RESET SILENCE button to continue, the
instrument will continue to beep and flash the LED arrays every few
seconds.
10. For each instrument, the SDM-GX Docking Station PC Controller Program will
indicate that an instrument was found and it will download the instrument’s
parameters. The BUMP
T
and CAL
S
LEDs will flash amber while the data is
being downloaded. Once the parameter download is complete, the BUMP
T
and CAL
S
LEDs will be solid amber. When all instruments have been turned
on, connected, and downloaded, the main program window will appear as
shown below. When the instrument is connected to the PC Controller Program,
FAIL
C--LIMIT
CAL
C--LIMIT