80 • Appendix A: Maintenance Mode
GasWatch 3 for CO and O
2
Entering Maintenance Mode
WARNING:
The GasWatch 3 is not in operation as a gas detector while in Maintenance
Mode.
1. Take the GasWatch 3 to a non-hazardous location and turn it off if it is on.
2. Press and hold AIR, then press and hold POWER/MODE. You will hear a beep after one
second. Continue to hold the buttons down.
3. When you hear a second beep, release the buttons.
ID DISP (pg.86)
ON
: User ID and Station ID screens appear in startup sequence. IDs can be changed in
Display Mode if
DISP.SET
in User Mode is also set to
ON
.
OFF
(factory setting): User ID and Station ID screens do not appear in startup sequence.
IDs cannot be changed in Display Mode.
ZERO.SUP
(pg.86)
ON
(factory setting): Not intended for field adjustment. The suppression value is 2 ppm.
ZERO.FLW
(pg.86)
ON
(factory setting): Not intended for field adjustment.
ZSUP.DSP (pg.86)
ON
: Zero suppression item appears in User Mode.
OFF
(factory setting): Zero suppression item does not appear in User Mode.
(Zero suppression item is always available in Maintenance Mode)
ZFLW.DSP
(pg.87)
ON
: Zero follower item appears in User Mode.
OFF
(factory setting): Zero follower item does not appear in User Mode.
(Zero follower item is always available in Maintenance Mode)
CYL.DISP (pg.87)
ON
(factory setting):
CYL SEL
item appears in
GAS CAL
.
OFF
:
CYL SEL
item does not appear in
GAS CAL
.
(RKI Instruments, Inc. does not recommend adjusting this setting.)
DATE (pg.87)
Set the current date and time.
PASS-W (pg.87)
ON
(factory setting): Maintenance Mode is password-protected. Factory-set password is
3000
.
OFF
: Maintenance Mode is not password-protected.
ROM/SUM
(pg.88)
View the firmware information for the GasWatch 3’s sensor board and main board.
Set all parameters back to their RKI factory settings.
START (pg.90)
Press and release POWER/MODE to begin the warmup sequence and enter Measuring
Mode.
Table 11: Maintenance Mode Menu Items
Maintenance
Mode
Menu Item
Description