90 • Appendix A: Maintenance Mode
GasWatch 3 for CO and O
2
4. Press and release POWER/MODE to perform a default operation. The instrument beeps
twice and returns to the
M.DEF
item in Maintenance Mode.
5. See “Entering Measuring Mode (START)” on page 90 to enter Measuring Mode.
Entering Measuring Mode (START)
1. While in Maintenance Mode, press AIR to scroll to
START
.
2. Press and release POWER/MODE. The instrument begins its warmup sequence.
START