GasWatch 3 for CO and O
2
Chapter 5: User Mode and Calibration • 59
1. After entering the
CAL SET
menu, press AIR to scroll to
CAL.EXPD
.
2. Press and release POWER/MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER/MODE to save the setting and return to the
CAL.EXPD
item
in the
CAL SET
menu.
5. See “Exiting the CAL SET Menu” on page 59 to return to User Mode.
Exiting the CAL SET Menu
1. While in the
CAL SET
menu, press AIR to scroll to
ESCAPE
.
2. Press and release POWER/MODE to return to the
CAL SET
item in User Mode.
3. See “Entering Measuring Mode (START)” on page 72 to enter Measuring Mode.
Setting Bump Test Parameters (BUMP.SET)
The
BUMP.SET
menu has 5 items:
SETTING
,
BP.RMDR
,
BP.INT
,
BP.EXPD
, and
ESCAPE
.
1. While in User Mode, press AIR to scroll to
BUMP.SET
.
2. Press and release POWER/MODE. The
SETTING
item appears.
SETTING
The
SETTING
menu has 5 items:
GAS.TIME
,
CHECK
,
CAL.TIME
,
A-CAL
, and
ESCAPE
.
1. After entering the
BUMP.SET
menu, press AIR to scroll to
SETTING
.
CAL .EXPD
BUMP .SET
SETTING
SETTING