GasWatch 3 for CO and O
2
Chapter 5: User Mode and Calibration • 57
8. Press and release POWER/MODE. The current setting flashes. In the example below, CO
is selected.
9. Use AIR to change the cylinder assignment. The choices are A, B, C, D, and E.
10. Press and release POWER/MODE to save the change. The cylinder assignment stops
flashing and the unit returns to the gas selection screen.
11. Repeat steps Step 7 - Step 10 for the other channel, if necessary.
12. Use AIR to scroll to
ESCAPE
.
13. Press and release POWER/MODE. The instrument returns to the
CYL SEL
menu item in
the Auto Cal Menu.
14. Press AIR to scroll to
ESCAPE
.
15. Press and release POWER/MODE. The instrument returns to the
A-CAL
menu item in the
GAS CAL
menu.
16. See “Exiting the GAS CAL Menu” on page 57 to return to User Mode.
Exiting the GAS CAL Menu
1. While in the
GAS CAL
menu, press AIR to scroll to
ESCAPE
.
2. Press and release POWER/MODE to return to the
GAS CAL
item in User Mode.
3. See “Entering Measuring Mode (START)” on page 72 to enter Measuring Mode.
Setting Calibration Parameters (CAL SET)
The
CAL SET
menu has 4 items:
CAL.RMDR
,
CAL.INT
,
CAL.EXPD
, and
ESCAPE
.
1. While in User Mode, press AIR to scroll to
CAL SET
.
2. Press and release POWER/MODE. The
CAL.RMDR
item appears.
CYL SEL
A
CO
ppm
CAL SET
CAL .RMDR