GasWatch 3 for CO and O
2
Chapter 5: User Mode and Calibration • 53
Option C from Flow Chart
Option D from Flow Chart
• Calibration passed
• One cylinder assigned
• Auto Start After Successful Cal set to
OFF (factory setting is ON)
• Calibration failed
1. The instrument indicates that all
channels passed the calibration.
2. The instrument shows the current gas
readings.
3. If
MAX.SPAN
is set to
ON
(default is
OFF
, see pg.93), the response reading’s
maximum adjustment is displayed. A
maximum span of 100 ppm on the CO
channel indicates that the reading could
have been adjusted up to 100 ppm. If
the maximum span value is close to the
calibration gas value, the sensor should
be replaced soon.
The maximum CO adjustment is either
twice the calibration value or full scale,
whichever is lower. The maximum
oxygen adjustment is 25.0% volume.
4. Close the regulator.
5. Unscrew the regulator.
6. Remove the calibration cup.
7. The instrument continues to the
START
item in the
E-CAL
menu.
8. Press and release POWER/MODE to
enter Measuring Mode.
1. The instrument indicates which
channels failed the calibration.
2. Close the regulator.
3. Unscrew the regulator.
4. Remove the calibration cup.
5. Press and release POWER/MODE to
return to the E-CAL CYL X screen in
the
E-CAL
menu.
6. See “Troubleshooting” on page 73.
E-CAL
PASS
%
PASS
ppm
E-CAL
0.5
%
51
ppm
MAX.SPAN
24.0
%
91
ppm
E-CAL
FAIL
%
FAIL
ppm