GasWatch 3 for CO and O
2
Appendix B: Gas Select Mode • 93
2. Press and release POWER/MODE. The display cycles through the following screens.
3. Press and release POWER/MODE to save the current alarm point settings as the default.
4. The instrument returns to the
SAVE-AP
item in Gas Select Mode.
5. See “Exiting Gas Select Mode (START)” on page 94 to enter Measuring Mode.
Turning Calibration Max Span On/Off (MAX.SPAN)
ON
: After a passed calibration, the GasWatch 3 displays the response reading’s maximum
adjustment. A maximum span of 100 ppm on the CO channel indicates that the reading could
have been adjusted up to 100 ppm. If the maximum span value is close to the calibration
value, the sensor should be replaced soon. The upper limit on the maximum CO adjustment is
either twice the calibration value or full scale, whichever is lower. The upper limit on the
maximum oxygen adjustment is 25.0% volume.
OFF
(factory setting): There is no maximum span indication at the end of a calibration.
1. While in Gas Select Mode, press AIR to scroll to
MAX.SPAN
.
2. Press and release POWER/MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER/MODE to save the setting and return to the
MAX.SPAN
item
in Gas Select Mode.
5. See “Exiting Gas Select Mode (START)” on page 94 to enter Measuring Mode.
SAVE-AP
SEL
YES .MODE
SEL
NO .AIR
SEL
MAX .SPAN