Chapter 5: User Mode and Calibration • 51
Setting Bump Test Parameters (BUMP.SET)
The
BUMP.SET
menu has 5 items:
SETTING
,
BP.RMDR
,
BP.INT
,
BP.EXPD
, and
ESCAPE
.
1. While in User Mode, press AIR to scroll to
BUMP.SET
.
2. Press and release POWER MODE. The
SETTING
item appears.
SETTING
The
SETTING
menu has 5 items:
GAS.TIME
,
CHECK
,
CAL.TIME
,
A-CAL
, and
ESCAPE
.
1. After entering the
BUMP.SET
menu, press AIR to scroll to
SETTING
.
2. Press and release POWER MODE. The
GAS.TIME
item appears.
GAS.TIME
The
GAS.TIME
is the amount of time that the instrument is exposed to gas during a bump
test. The available choices are
30
seconds (factory setting),
45
seconds,
60
seconds, and
90
seconds.
1. After entering the
SETTING
menu, press AIR to scroll to
GAS.TIME
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
GAS TIME
item in
the
SETTING
menu.
5. See “Exiting the SETTING Menu” on page 53 to return to
BUMP.SET
menu.
BUMP .SET
SETTING
SETTING
GAS .TIME