SC-04
Appendix A: Maintenance Mode • 81
Turning the Auto Zero Function On/Off (A.ZERO)
ON
: The SC-04 asks if you want to perform a fresh air adjustment at the end of the startup
sequence.
OFF
(factory setting): The SC-04 does not ask if you want to perform a fresh air adjustment at
the end of the startup sequence.
1. While in Maintenance Mode, press AIR to scroll to
A.ZERO
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
A.ZERO
item in
Maintenance Mode.
5. See “Entering Measuring Mode (START)” on page 85 to enter Measuring Mode.
Turning the ID Display Function On/Off (ID DISP)
ON
: The User ID and Station ID screens appear in the startup sequence. If
DISP.SET
in User
Mode is also set to
ON
, then the IDs can be changed in Display Mode.
OFF
(factory setting): The User ID and Station ID screens do not appear in the startup
sequence and the IDs cannot be changed in Display Mode.
1. While in Maintenance Mode, press AIR to scroll to
ID DISP
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
ID DISP
item in
Maintenance Mode.
5. See “Entering Measuring Mode (START)” on page 85 to enter Measuring Mode.
A .ZERO
ID DISP