40 • Chapter 5: User Mode and Calibration
SC-04
Option C from Flow Chart
Option D from Flow Chart
• Bump test failed
• A-CAL set to ON (factory setting)
• Calibration passed
• Bump test failed
• A-CAL set to ON (factory setting)
• Calibration failed
1. A calibration immediately and
automatically starts. Continue to apply
the calibration gas.
2. The calibration time is the difference
between the
GAS.TIME
and the
CAL.TIME
values defined in the
BUMP.SET\SETTING
item in User
Mode.
3. The instrument shows the pass/fail
results of the bump test/calibration. Use
AIR to scroll between the results and
the readings.
4. Close the regulator.
5. Unscrew the regulator.
6. Remove the calibration cup.
7. Press and release POWER MODE to
return to the
BUMP
screen in the
BUMP
menu.
8. Use AIR to scroll to
START
and press
and release POWER MODE to enter
Measuring Mode.
1. A calibration immediately and
automatically starts. Continue to apply
the calibration gas.
2. The calibration time is the difference
between the
GAS.TIME
and the
CAL.TIME
values defined in the
BUMP.SET\SETTING
item in User
Mode.
3. The instrument shows the pass/fail
results of the bump test/calibration. The
LED flashes and the buzzer sounds. Use
AIR to scroll between the results and
the readings.
4. Close the regulator.
5. Unscrew the regulator.
6. Remove the calibration cup.
7. Press and release POWER MODE to
return to the
BUMP
screen in the
BUMP
menu.
8. Use AIR to scroll to
START
and press
and release POWER MODE to enter
Measuring Mode.
SO2
0.00
CAL 60
ppm
SO2
0.00
APLY 60
ppm
SO2
F P
BMP / CAL
ppm
SO2
0.90
BMP /
ppm
SO2
4.70
/ CAL
ppm
AIR
AIR
SO2
0.00
CAL 60
ppm
SO2
0.00
APLY 60
ppm
SO2
F F
BMP / CAL
ppm
SO2
0.00
BMP /
ppm
SO2
0.90
/ CAL
ppm
AIR
AIR