SC-04
Chapter 1: Introduction • 7
Specifications
Table 1: Standard Sensor Specifications/Alarm Points
Table 2: SC-04 Specifications
Ammonia
(NH
3
)
Chlorine
(Cl
2
)
Hydrogen
Cyanide
(HCN)
Nitrogen
Dioxide
(NO
2
)
Phosphine
(PH
3
)
Sulfur
Dioxide
(SO
2
)
Detection Range
0 - 400.0 ppm
0 - 20.00
ppm
0 - 30.0
ppm
0 - 20.00
ppm
0 - 20.00
ppm
0 - 100.00
ppm
Lowest Detectable
Limit (LDL)
4.0 ppm
0.10 ppm
1.0 ppm
0.30 ppm
0.02 ppm
0.20 ppm
Reading Increment
0.5 ppm
0.05 ppm
0.1 ppm
0.05 ppm
0.01 ppm
0.05 ppm
Warning Setpoint
25 ppm
1.00 ppm
5.0 ppm
2.00 ppm
0.30 ppm
2.00 ppm
Alarm Setpoint
35 ppm
2.00 ppm
10.0 ppm
4.00 ppm
0.60 ppm
5.00 ppm
Alarm H Setpoint
300 ppm
10.00 ppm
30.0 ppm
20.00 ppm
0.60 ppm
100.00 ppm
STEL Setpoint
35 ppm
1 ppm
10.0 ppm
1.00 ppm
1.00 ppm
5.00 ppm
TWA Setpoint
25 ppm
0.5 ppm
4.7 ppm
0.50 ppm
0.30 ppm
2.00 ppm
Sampling
Method
Diffusion
Response Time
T90 Within 10 Seconds
Display
Graphics LCD Display
Operating
Temperature
& Humidity
Continuous environment
: -20°C to 50
°
C/10 to 90% RH
Temporary environment (up to 15 minutes)
: -40°C to 60
°
C/0 to 95% RH
Indication
Accuracy
± 10% of reading or ± 5% of full scale (whichever is greater)
Safety/
Regulatory
• ATEX: Certificate Number DEKRA 19ATEX0097
II 1G Ex ia IIC T4 Ga (with alkaline batteries)
II 1G Ex ia IIC T3, Ga (with Ni-MH batteries)
• IECEx: Certificate Number IECEx DEK 19.0059
Ex ia IIC T4 Ga (with alkaline batteries)
Ex ia IIC T3 Ga (with Ni-MH batteries)
• QPS classified, “C/US”, as Intrinsically Safe. Exia. Class I, Groups A, B, C, & D.
Instrument
Power
Information
• Operating Voltage: 3.0V
• Operating Current: 1.0 mA
• Operating Power: 3.0 mW
Power Supply
2 AAA alkaline batteries; 1.5V, 1.175 AH (Duracell MN2400 or PC2400)
OR
2 AAA Ni-MH batteries; 1.2V, 800 mAH (Panasonic Eneloop BK-4MCC)