Chapter 5: User Mode and Calibration • 53
4. Press and release POWER MODE to save the setting and return to the
A-CAL
item in the
SETTING
menu.
5. See “Exiting the SETTING Menu” on page 53 to return to
BUMP.SET
menu.
Exiting the SETTING Menu
1. While in the
SETTING
menu, press AIR to scroll to
ESCAPE
.
2. Press and release POWER MODE to return to the
SETTING
item in the
BUMP.SET
menu.
3. See “Exiting the BUMP.SET Menu” on page 54 to return to User Mode.
BP.RMDR
ON
: The SC-04 gives an indication at start up if it is due for bump testing. The type of
indication depends on the
BP.EXPD
setting (see pg.54). If the instrument is not due for bump
testing, a check mark appears in the lower left corner of the LCD.
OFF
(factory setting): The SC-04 does not give an indication at start up if it is due for bump
testing.
1. After entering the
BUMP.SET
menu, press AIR to scroll to
BP.RMDR
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
BP.RMDR
item in
the
BUMP.SET
menu.
5. See “Exiting the BUMP.SET Menu” on page 54 to return to User Mode.
BP.INT
This setting defines the amount of time between bump tests and can be set in 1 day
increments. The minimum setting is
0
days and the maximum setting is
30
days (factory
setting).
1. After entering the
BUMP.SET
menu, press AIR to scroll to
BP.INT
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
BP.INT
item in the
BUMP.SET
menu.
5. See “Exiting the BUMP.SET Menu” on page 54 to return to User Mode.
BP .RMDR
BP .INT