54 • Chapter 5: User Mode and Calibration
SC-04
BP.EXPD
This item defines what indication is given during start up when a bump test is due and
BP.RMDR
is set to
ON
(factory setting is
OFF
).
CONFIRM
(factory setting): The SC-04 gives an indication at start up if a bump test is past
due. Press and release AIR to continue without bump testing or press and release POWER
MODE to enter User Mode and perform a bump test.
CANT.USE
: The SC-04 gives an indication at start up that a bump test is past due. Press and
release POWER MODE to enter User Mode and perform a bump test. Pressing AIR has no
effect. A successful bump test must be performed in order to use the instrument.
NONE
: The SC-04 gives an indication at startup that a bump test is past due. It is not
necessary to acknowledge the indication. If desired, press POWER MODE to enter User
Mode and perform a bump test. The warm-up sequence continues on its own.
1. After entering the
BUMP.SET
menu, press AIR to scroll to
BP.EXPD
.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the
BP.EXPD
item in
the
BUMP.SET
menu.
5. See “Exiting the BUMP.SET Menu” on page 54 to return to User Mode.
Exiting the BUMP.SET Menu
1. While in the
BUMP.SET
menu, press AIR to scroll to
ESCAPE
.
2. Press and release POWER MODE to return to the
BUMP.SET
item in User Mode.
3. See “Entering Measuring Mode (START)” on page 62 to enter Measuring Mode.
Alarm Settings (ALARM--P)
The
ALARM--P
menu has 3 items:
ALARM--P
,
DEF.ALMP
, and
ESCAPE
.
1. While in User Mode, press AIR to scroll to
ALARM--P
.
2. Press and release POWER MODE. The first screen displays.
BP .EXPD
ALARM-P
SO2
ALARM-P
ppm