Reference Manual Rev 1.13
Page 90
004R-646-113
13.4. Calibration Errors
Following are a list of the possible error messages that may be displayed to warn of
failed or incorrect calibration:
Error
Description
Resolution
(FAILED)
(BAND)
An attempt has been made to
calibrate with a weight or
signal which is not in the valid
range.
Check weights and retry.
(FAILED)
(ERROR)
An attempt has been made to
calibrate while the scale
signal is not valid.
Check loadcell connection and
the 4-wire/6-wire setting.
(FAILED)
(TIMEOUT)
For an unknown reason, the
calibration was unable to
complete.
Retry.
(FAILED)
(RES)
An attempt has been made to
calibrate the scale to a
resolution which is too high
for the instrument.
Check weights and retry.
(FAILED)
(TOO CLOSE)
An attempt has been made to
add a linearisation point too
close to zero, span or another
linearisation point.
Check weights and retry.
Table 5: Calibration errors
Содержание AxleweighR R420
Страница 45: ...Reference Manual Rev 1 13 004R 646 113 Page 45 8 6 Two Pass 8 7 Two Pass Automatic Temporary ID Allocated...
Страница 48: ...Reference Manual Rev 1 13 Page 48 004R 646 113 9 5 Two Pass 9 6 Two Pass Automatic Temporary ID Allocated...
Страница 51: ...Reference Manual Rev 1 13 004R 646 113 Page 51 10 5 Two Pass 10 6 Two Pass Automatic Temporary ID Allocated...
Страница 85: ...Reference Manual Rev 1 13 004R 646 113 Page 85 13 1 1 ZERO Zero Calibration Routine...