Reference Manual Rev 1.13
Page 128
004R-646-113
Name
Address Type
Description
Response (ok): 81100103:00000000
Send (status?): 20110021
Response (calibrating): 81110021:0000A400
Send (status?): 20110021
Response (calibrating): 81110021:0000A400
Send (status?): 20110021
Response (not calibrating): 81110021:00008400
Clear Linearity *
0105
H
261
d
Execute
This register clears a previously entered
linearisation calibration. There are 10
linearisation points [numbered 0 .. 9] which
can be cleared separately. The linearisation
point to clear is sent as a parameter.
Example:
Sent (Clear 1st point): 20100105:0
Response (ok): 81100105:00000000
Direct zero calibration*
0106
H
262
d
Execute
This register is used to perform a direct zero
calibration in the same way as the direct zero
calibration via the menus. A direct zero
calibration is very fast and the display may
not change in the same way as a zero
calibration.
The mV/V value is sent as a parameter. It is
sent as mV/V x 10000. Example:
• 0.5mV/V → 5000 → 1388H
• 1.0mV/V → 10000 → 2710H
• 2.5mV/V → 25000 → 61A8H
Example:
Sent (0.5mV/V): 20100106:1388
Response(ok): 81100106:00000000
Direct span calibration*
0107
H
263
d
Execute
This register is used to perform a direct span
calibration in the same way as the direct span
calibration via the menus. A direct span
calibration is very fast and the display may
not change in the same way as a span
calibration.
The mV/V value OF FULLSCALE is sent as a
parameter. It is sent as mV/V x 10000. E.g:
• 0.5mV/V → 5000 → 1388H
• 1.0mV/V → 10000 → 2710H
• 2.5mV/V → 25000 → 61A8H
Example use:
Sent (1.0mV/V): 20100107:2710
Response(ok): 81100106:00000000
Gross Weight
0149
H
329
d
Number
Gross weight in counts
Net Weight
014A
H
330
d
Number
Net weight in counts
Current Time/Date
0150
H
String
Read this register to get instrument date/time
Содержание AxleweighR R420
Страница 45: ...Reference Manual Rev 1 13 004R 646 113 Page 45 8 6 Two Pass 8 7 Two Pass Automatic Temporary ID Allocated...
Страница 48: ...Reference Manual Rev 1 13 Page 48 004R 646 113 9 5 Two Pass 9 6 Two Pass Automatic Temporary ID Allocated...
Страница 51: ...Reference Manual Rev 1 13 004R 646 113 Page 51 10 5 Two Pass 10 6 Two Pass Automatic Temporary ID Allocated...
Страница 85: ...Reference Manual Rev 1 13 004R 646 113 Page 85 13 1 1 ZERO Zero Calibration Routine...