RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-29
Key Points
Y-T:
The conventional oscilloscope display format. It shows the voltage of a waveform
record (on the vertical axis) as it varies over time (on the horizontal axis).
X-Y:
X-Y format displays channel 1 in the horizontal axis and channel 2 in the vertical axis.
Roll Mode:
In this mode, the waveform display rolls from right to left. No trigger or horizontal
offset control of waveforms is available during Roll Mode, and it’s only available
when set to 500 ms/div or slower.
Slow Scan Mode:
This mode is available when the horizontal time base is set to 50ms or slower. In this
mode, the oscilloscope acquires sufficient data for the left part to the trigger point,
then wait for trigger, when trigger occurs, it continue to draw the rest part from the
trigger point to the end of the right side. When choosing this mode to view low
frequency signals, it is recommended that the channel coupling be set as DC.
Time/Div:
Horizontal scale. If the waveform acquisition is stopped (using the
button), the
Time/Div control expands or compresses the waveform.
Содержание VS5000 Series
Страница 2: ......
Страница 37: ...RIGOL 2008 RIGOL Technologies Inc User s Guide for VS5000 Series 2 5 Figure 2 3 DC Coupling Setting DC Coupling...
Страница 70: ...2008 RIGOL Technologies Inc User s Guide for VS5000 Series 2 38 Figure 2 24 Video trigger Line Synchronization...
Страница 83: ...RIGOL 2008 RIGOL Technologies Inc User s Guide for VS5000 Series 2 51 Figure 2 31 Peak Detect...
Страница 114: ......
Страница 118: ......
Страница 129: ...RIGOL 2008 RIGOL Technologies Inc User s Guide for VS5000 Series 3 11 Figure 3 10 FFT Measurement Cursor Type X...
Страница 131: ...RIGOL 2008 RIGOL Technologies Inc User s Guide for VS5000 Series 3 13 Figure 3 11 Pass Fail Test...