Example 3: Capture a Single-Shot Signal
To capture a single event, it needs to gather some pre-test knowledge of the signal in
order to set up the trigger level and slope correctly. For example, if the event is
derived from TTL logic, a trigger level of 2 Volts should work on a rising edge.
Do these steps as follows:
1
.
Set the probe and the channel attenuations to 1:10.
2
.
Set up the trigger.
1)
Click
to show
Trigger
window.
2)
Adjust the trigger Mode to
Edge
. Set the Source to
CH1
and the Slop to
Rising
. Set the Sweep type to
Single
.
3)
Adjust the Volts/Div and the time base in a proper range for the signal
4)
Set the trigger in the
Setting
window of the
Trigger
window, or drag the
trigger level sign on the waveform display screen.
5)
Click
to start capturing. When the trigger conditions are met, data
appears on the display representing the data points that the oscilloscope
obtained with one acquisition.
This function helps to capture the signal occurrence easily, such as the noise with
large amplitude; set the trigger level higher a little above the normal level and press
and wait. When noise occurs, the instrument will record the waveform before
and after the trigger.
Содержание VS5000 Series
Страница 2: ......
Страница 37: ...RIGOL 2008 RIGOL Technologies Inc User s Guide for VS5000 Series 2 5 Figure 2 3 DC Coupling Setting DC Coupling...
Страница 70: ...2008 RIGOL Technologies Inc User s Guide for VS5000 Series 2 38 Figure 2 24 Video trigger Line Synchronization...
Страница 83: ...RIGOL 2008 RIGOL Technologies Inc User s Guide for VS5000 Series 2 51 Figure 2 31 Peak Detect...
Страница 114: ......
Страница 118: ......
Страница 129: ...RIGOL 2008 RIGOL Technologies Inc User s Guide for VS5000 Series 3 11 Figure 3 10 FFT Measurement Cursor Type X...
Страница 131: ...RIGOL 2008 RIGOL Technologies Inc User s Guide for VS5000 Series 3 13 Figure 3 11 Pass Fail Test...